Mini Review
Frontiers in Anesthesiology ·
Volume 5 - 2026 | https://doi.org/10.3389/fanes.2026.1948602
Closed-loop non-invasive ventilation in critical care: current evidence and clinical perspectives
- Fernando S. Guimaraes1*corresponding author
- Antonio M. Esquinas2
- 1Federal University of Rio de Janeiro, Rio de Janeiro, Brazil
- 2Intensive Care Unit, Hospital Morales Meseguer NIV-ICU Group, Biomedical Research Institute Pascual Parrilla-IMIB, Murcia, Spain
Abstract
Non-invasive ventilation (NIV) is a cornerstone of acute and perioperative respiratory support, but its effectiveness is constrained by patient selection, interface leaks, patient-ventilator asynchrony, excessive inspiratory effort, and delayed recognition of treatment failure. Closed-loop and adaptive approaches may address some of these limitations by allowing ventilators or decision-support systems to adjust support according to physiological feedback. This mini review synthesizes current evidence on closed-loop concepts relevant to NIV in critical care, including leak-compensation algorithms, automated triggering and cycling, waveform-guided synchronization, proportional and neurally adjusted ventilatory assistance, volume-assured pressure support, monitoring of respiratory drive and effort, and emerging artificial intelligence approaches. The available literature suggests that most clinically available NIV technologies should be regarded as partial closed-loop systems rather than fully autonomous physiological controllers. Evidence is strongest for improved patient-ventilator interaction through leak-adapted algorithms, dedicated NIV ventilators, NAVA, and automated waveform-guided cycling, whereas direct evidence that closed-loop NIV improves patient-centred outcomes in critically ill adults remains limited. Physiological monitoring, particularly early assessment of inspiratory effort, may provide key input signals for future closed-loop NIV strategies designed to personalize support and avoid delayed intubation. Artificial intelligence, ontologies, digital twins, and reinforcement learning may accelerate development, but their current role is mainly preclinical or conceptual. Future trials should test clinically supervised, safety-constrained systems that integrate ventilator waveforms, gas exchange, leaks, comfort, drive, and failure prediction.
Keywords
Introduction
Non-invasive ventilation (NIV) has become an essential component of respiratory support in acute, critical care, and perioperative settings. Guidelines support NIV for acute hypercapnic respiratory failure, cardiogenic pulmonary edema, and selected post-extubation or postoperative scenarios, whereas its use in de novo hypoxemic respiratory failure requires careful patient selection and close monitoring (1, 2). The Research Topic “Critical Care Focus in Non-Invasive Ventilatory Approaches: a Loop Approaches” emphasizes technological advances, monitoring, hypercapnic respiratory failure, critical care anesthesiology, and continuity of care across phases of treatment. These priorities make closed-loop and adaptive NIV an attractive, but still underdeveloped, field.
Table 1. Comparative analysis of current and emerging closed-loop strategies relevant to non-invasive ventilation in critical care.
| System/strategy (loop type) | Feedback/target | Automated output | Evidence/clinical benefit | Main limitation |
|---|---|---|---|---|
| Leak-adapted NIV/automated trigger-cycling (partial feedback loop; vendor-specific adaptive algorithm) | Flow, pressure, leak estimate; timing targets | Trigger/cycling thresholds, flow estimation | Bench and clinical NIV studies support improved but device-dependent synchrony; current waveform automation is promising (3–5, 20–23) | Performance varies across devices/interfaces; correcting one asynchrony may create another |
| NAVA (servo) | EAdi; neural timing/drive | Neural triggering/cycling and assistance proportional to EAdi | Physiological studies improve synchrony/comfort; RCTs reduce asynchrony but have not consistently reduced NIV failure (29–32) | Special catheter/equipment; clinical-outcome benefit unproven |
| PAV/PAV + (servo) | Estimated effort and respiratory mechanics | Assistance proportional to patient effort | Strong physiological rationale; evidence mainly from invasive assisted ventilation (11) | Limited acute-care NIV evidence |
| AVAPS/iVAPS (adaptive) | Tidal-volume or alveolar-ventilation target | Inspiratory pressure within preset limits; selected systems can auto-adjust EPAP | Clinically available; AVAPS meta-analysis suggests lower intubation in selected acute hypercapnic populations (34–38) | Does not directly monitor effort, synchrony, or failure trajectory; evidence heterogeneous |
| ASV/ASV with IntelliSync (adaptive/optimal targeting) | Respiratory mechanics/ventilation target; waveform-derived timing with IntelliSync | Automated respiratory pattern and support; automated trigger/cycling with IntelliSync | NIV RCTs in AECOPD show similar failure for NAVA vs ASV and feasibility signals for ASV-IntelliSync (32, 33) | NIV outcome evidence remains limited; device-specific implementation |
| INTELLiVENT-ASV (multivariable closed loop) | Ventilation plus gas-exchange feedback | Automated ventilation and oxygenation-related settings within protocolized targets | Large invasive ACTiVE trial improved ventilation quality but not ventilator-free days (19) | Direct critical-care NIV evidence is lacking |
| SmartCare/PS (intelligent automated weaning) | Respiratory rate, tidal volume and CO₂-related variables within predefined comfort zones | Automated pressure-support adjustments and progression toward separation | Established concept for automated invasive weaning (11) | Not designed for NIV-specific leak handling or NIV-failure decisions |
| Optimal-targeting concepts (ASV, AVM2, MFV) | Mathematical objective/cost function | Selected respiratory-rate and tidal-volume combination | Different objective functions produce different “optimal” VT/rate combinations; simulated VT ≈7.4, 6.2, and 6.7 mL/kg for ASV, AVM2, and MFV, respectively (12) | Model assumptions and chosen objective determine the solution |
| AI-enabled decision support/future intelligent loops | Waveforms, EHR, gas exchange, imaging, device metadata; temporally defined features | Phenotyping, failure prediction, safety-constrained recommendations or future control policies | Ontology now peer-reviewed; AI methodology and digital-twin/RL work remain largely preclinical or non-NIV-specific (39–41) | Requires reproducible temporal feature design, calibration, external validation, explainability, and clinician oversight |
The clinical promise of NIV is inseparable from its technical fragility. Unlike invasive ventilation, NIV operates through a non-hermetic interface, with intentional and unintentional leaks that can distort flow estimation, trigger performance, cycling-off, delivered tidal volume, and displayed monitoring variables (3–5). NIV also preserves spontaneous breathing. This can maintain diaphragm activity and avoid endotracheal intubation, but may expose the lungs and respiratory muscles to injurious effort if support is insufficient or if treatment failure is recognized too late (6–8). In this context, the ideal NIV system would not merely deliver a prescribed pressure; it would continuously sense leaks, respiratory timing, mechanics, gas exchange, inspiratory effort, and clinical trajectory, then adjust support while preserving clinician oversight.
This mini-review focuses on closed-loop and adaptive strategies for NIV in critical care. Because fully autonomous NIV systems remain uncommon, the review adopts a pragmatic definition: technologies are considered relevant when they use physiological or ventilator-derived feedback to modify ventilatory assistance, trigger/cycling behavior, delivered pressure, volume targeting, or clinical decision support. In control terms, a closed loop requires a reference or target, a measured or estimated output that is fed back to the controller, and an actuator that changes ventilatory assistance (9). Monitoring variables are therefore feedback elements, whereas clinical context and clinician-defined goals provide supervisory inputs. The central argument is that current NIV practice already contains several partial loops, but these components have not yet been integrated into robust, outcome-tested closed-loop NIV platforms. Figure 1 summarizes this architecture and its safety constraints.

What should count as closed-loop NIV?
Closed-loop ventilation is best understood as a control architecture rather than as a single proprietary ventilator mode. Chatburn and Mireles-Cabodevila classified ventilator targeting schemes as set-point, dual, servo, adaptive, optimal, and intelligent according to how preset goals are achieved (10). Lellouche and Brochard emphasized that PAV, NAVA, ASV, and SmartCare represent advanced closed loops with different inputs, response times, and clinical objectives (11). Van der Staay and Chatburn further showed that “optimal” targeting depends on the objective function: at comparable simulated ventilation efficiency in adult ARDS, expected tidal volume was approximately 7.4 mL/kg with ASV, 6.2 mL/kg with AVM2, and 6.7 mL/kg with MFV, with different optimal frequencies generated by the different cost functions (12). Thus, automation embeds a selected physiological objective rather than eliminating clinical judgment.
For NIV, this taxonomy matters because the term “closed-loop NIV” can be overused. AVAPS and iVAPS adapt inspiratory pressure to a ventilation target but do not monitor effort or failure risk; leak-compensation algorithms modify triggering and flow estimation without choosing clinical goals; and NAVA provides servo assistance proportional to EAdi without deciding whether NIV remains appropriate. A mature system would probably be hierarchical and multirate: fast breath-to-breath loops would address leak, triggering, and cycling, whereas slower loops would target ventilation, effort, gas exchange, and failure risk. Stability requires accounting for sensor latency, physiological delays, signal noise, inter- and intra-patient variability, competing targets, and nonlinear or path-dependent behavior. In the respiratory system, pressure-volume hysteresis can reflect recruitment/derecruitment and viscoelastic behavior, illustrating that the response to a given pressure may depend on the preceding state (13). In physiological closed-loop devices, sensor latency and dynamic uncertainty can impair controller stability, supporting bounded rates of change, fail-safe or fallback modes, and clinician override (9, 14). Feedback inputs should distinguish directly measured device signals (airway pressure, device flow), model-derived estimates (leak, patient flow, tidal volume, mechanics), and additional physiological measurements or surrogates (EAdi, esophageal pressure, P0.1, SpO2, ultrasound), which differ in measurement pathway, susceptibility to artifact or estimation error, and temporal resolution (14–17). These categories should not be interpreted as an ordinal reliability scale; signal validity is device-, interface-, and context-dependent.
The evidence base in invasive ventilation shows both the potential and the limits of this concept. A systematic review found that closed-loop ventilation may improve process measures and reduce staff workload, although the effects on clinical outcomes vary across systems and populations (18). The large ACTiVE trial reported that early INTELLiVENT-ASV did not increase ventilator-free days compared with protocolized conventional invasive ventilation, despite improved ventilation quality and fewer severe gas-exchange derangements (19). These findings are relevant to NIV because they caution against assuming that better automation will necessarily translate into patient-centered benefit.
Leak compensation, triggering, and cycling: the first practical NIV loops
The most concrete closed-loop problem in NIV is the leak. In a bench model, Vignaux and colleagues demonstrated that leaks impair ICU ventilator functions by increasing trigger delay and triggering workload, reducing pressurization, and delaying cycling; NIV-specific modes partially corrected these effects, but performance varied widely, and some ventilators worsened leak-related dysfunction (3). In a multicenter clinical study, severe asynchrony during NIV for acute respiratory failure was common: an asynchrony index greater than 10% occurred in 43% of patients, and leak magnitude correlated with ineffective efforts and delayed cycling (20). A subsequent randomized crossover clinical study showed that NIV algorithms reduced leak-related asynchronies, but global asynchrony did not uniformly improve because some algorithms increased premature cycling (21). These studies define a core principle for closed-loop NIV: an algorithm that corrects one form of asynchrony can create another if it is not physiologically integrated.
Carteaux and colleagues extended this point by comparing ICU, transport, and dedicated NIV ventilators in bench and clinical settings. Dedicated NIV ventilators achieved lower asynchrony indices than ICU ventilators with or without their NIV algorithms, largely because of better handling of leaks and auto-triggering (4). Recent bench evidence and a 2026 review of patient-ventilator interaction confirm that modern devices still vary in leak handling and that unintentional leak remains a major determinant of trigger and cycle discordance (5, 22, 23). Trigger and cycling asynchronies, including ineffective efforts, auto-triggering, and delayed or premature cycling, should therefore be interpreted together with leak magnitude, pressure/flow waveform morphology, respiratory mechanics, and support level rather than by a global asynchrony index alone (20–25). Thus, accurate estimation of patient flow despite intentional and unintentional leaks remains a prerequisite for advanced NIV control (15).
Cycling-off is particularly important. Calderini and colleagues showed that, in the presence of mask leaks, time-cycled inspiratory pressure support improved patient-machine interaction compared with conventional flow-cycled NIV in severe hypoxemic respiratory failure (26). Tassaux and colleagues later showed in intubated patients with obstructive lung disease that increasing the expiratory trigger threshold reduced delayed cycling, intrinsic PEEP, non-triggering breaths, and inspiratory muscle workload (27). Although this latter study was not performed during NIV, its physiological mechanism is directly relevant to NIV algorithms in COPD and other obstructive conditions. Chen and colleagues tested automated cycling during noninvasive pressure support ventilation in a lung simulator and found better synchronization with Auto-Trak than with conventional cycling approaches across simulated mechanics and leaks (28).
Waveform-guided synchronization and proportional assistance
A key step toward clinically useful closed-loop NIV is the ability to infer the timing of patient effort from ventilator waveforms without invasive monitoring. Mojoli and colleagues developed a waveform method using airway pressure and flow, with esophageal pressure as the reference, that accurately detected respiratory efforts, trigger delay, cycling delay, and early cycling (24). The same group then tested automated real-time waveform-guided cycling-off in 15 difficult-to-wean patients. Automation reduced cycling delay, ineffective efforts, and total asynchrony time compared with standard cycling. It also performed as well as or better than protocol-defined expert adjustment of expiratory trigger sensitivity, but the restricted expert intervention precludes interpreting this as superiority over comprehensive expert ventilator management (25). Although conducted during invasive pressure support, these studies provide a mechanistic bridge to NIV, where continuous noninvasive effort-timing detection would be particularly valuable.
Direct NIV evidence is also emerging. Renaud and colleagues evaluated an automated real-time waveform analysis algorithm during NIV in a bench model and reported improved patient-ventilator synchrony under leak conditions (22). This study is highly relevant because it tests a technological solution tailored to variable leaks that distort pressure and flow signals. However, it used a single ventilator, calibrated circuit leaks without a face-mask interface, and three predefined simulated respiratory-mechanics conditions; therefore, the evidence remains preclinical and cannot establish performance across devices or clinical NIV conditions. Clinical trials are still needed to determine whether automated waveform-guided NIV reduces discomfort, intubation, sedation use, or time under ineffective support.
NAVA represents a different strategy: instead of inferring neural timing from pneumatic waveforms, it uses EAdi to trigger, cycle, and proportionally assist each breath. Piquilloud and colleagues showed that NAVA improved patient-ventilator interaction during NIV delivered by face mask compared with conventional pressure support (29). Longhini and colleagues later showed that optimized NIV-NAVA settings could improve comfort and synchrony compared with pneumatic triggering and cycling (30). Randomized clinical evidence tempers these physiological findings: in adults with de novo acute respiratory failure, NAVA reduced asynchrony and NIV-related complications but not NIV failure or 28-day mortality compared with pressure support ventilation (31). In AECOPD, NAVA and ASV had similar NIV failure, asynchrony, and 90-day mortality, although ASV required fewer clinician adjustments (32). A 2026 feasibility trial of ASV with IntelliSync versus pressure support ventilation in AECOPD found greater comfort and shorter hospital stay, with a numerically lower but nonsignificant NIV-failure rate; larger trials are required (33). PAV and PAV + remain conceptually relevant because they adjust support in proportion to patient effort and mechanics, but evidence in critical care NIV is less developed than in invasive assisted ventilation (11).
Volume-assured and adaptive pressure modes
Volume-assured pressure support modes are not full closed-loop NIV systems, but they represent a clinically available form of adaptive NIV. AVAPS adjusts inspiratory pressure support within a prescribed range to maintain a target tidal volume, while iVAPS targets alveolar ventilation and may incorporate auto-EPAP in selected devices. Shah and colleagues summarized these modes as extensions of bilevel NIV intended to stabilize ventilation when mechanics or effort change over time (34). In acute hypercapnic respiratory failure, Briones Claudett and colleagues reported clinical experience with BiPAP S/T-AVAPS, supporting the plausibility of adaptive pressure targeting in the acute setting (35). In chronic hypercapnic COPD, AVAPS and iVAPS studies suggest potential benefits in gas exchange, symptoms, and quality of life, although populations, devices, and treatment settings differ substantially (36, 37).
The most recent synthesis on this topic is the 2026 systematic review and meta-analysis by Gregory and colleagues. Across 11 comparative studies involving adults with acute hypercapnic respiratory failure or similar acute-care indications, AVAPS was associated with a lower risk of intubation than BiPAP, without a clear short-term mortality benefit and with similar physiological outcomes and hospital length of stay (38). However, only three studies contributed complete intubation data; the evidence mixed randomized and observational designs, was dominated by COPD populations, used variable outcome time points, and carried low-to-moderate risk-of-bias concerns. The pooled finding should therefore be interpreted as evidence for AVAPS in selected acute hypercapnic populations rather than evidence that closed-loop NIV as a class improves clinical outcomes. AVAPS remains a pressure-adaptation loop targeting volume and does not directly measure patient effort, dyspnea, synchrony, or progression toward failure.
Effort, drive, and failure prediction as missing loop inputs
A major limitation of current NIV automation is that most systems respond to pressure, flow, volume, and leaks, but not to the patient's injurious effort or probability of NIV failure. This gap is clinically important. Noninvasive respiratory support can reduce the work of breathing, but persistent high respiratory drive and effort may promote respiratory muscle fatigue, cardiovascular compromise, oxygen-delivery imbalance, and potentially patient self-inflicted lung injury (6, 8). The same review literature emphasizes that delayed intubation remains a central mechanism of harm when noninvasive support fails (2, 6).
Inspiratory effort monitoring may therefore become a key input for closed-loop NIV. Tonelli and colleagues studied patients with de novo hypoxemic respiratory failure undergoing NIV and found that relief of inspiratory effort, assessed by esophageal pressure within the first 2 h, was an early and accurate predictor of NIV outcome (7). This study is especially relevant because it connects physiological monitoring to a clinically actionable decision: whether NIV is unloading the patient sufficiently or merely postponing intubation. Telias and Spadaro reviewed methods to monitor respiratory drive and effort, including esophageal pressure, airway occlusion pressure, EAdi, and ultrasound, while emphasizing the need to interpret these signals in relation to respiratory muscle function and ventilator settings (16). Cammarota and colleagues further positioned esophageal pressure, EAdi, electrical impedance tomography, lung ultrasound, and respiratory muscle ultrasound as advanced bedside tools for personalized protection of the lung and respiratory muscles during acute respiratory failure (17).
Future NIV controllers should therefore move from synchrony loops toward injury-avoidance loops. Such systems would not simply minimize trigger delay or achieve a target tidal volume. They would aim to maintain adequate unloading while avoiding excessive tidal volume, excessive negative pleural pressure swings, high respiratory rate, worsening gas exchange, and poor comfort. The clinical output may sometimes not be a ventilator adjustment, but a recommendation to change interface, increase monitoring intensity, or proceed to intubation.
Artificial intelligence and future closed-loop NIV
Artificial intelligence (AI) may help integrate heterogeneous data streams that are difficult to combine through rule-based control alone. However, model performance depends on how variables and time are represented. Across ventilator-waveform and electronic health record applications, feature selection, sampling frequency, aggregation windows, temporal alignment, prediction horizon, missing-data handling, label definition, and class balance determine what physiology a model can learn; a breath-by-breath signal and a 30-minute trend are not interchangeable inputs. A systematic methodological review of longitudinal electronic health record prediction models highlights the importance of these design choices for temporally structured clinical data (39). Current AI work in ventilation remains largely retrospective, simulation-based, or invasive rather than NIV-specific, so the near-term opportunities are signal interpretation, phenotyping, failure prediction, and safety testing rather than autonomous control.
Three lines of work are particularly relevant. Ontology-based representations may standardize NIV concepts across electronic health records, devices, indications, parameters, and outcomes; Islam, Mosier, and Subbian developed and evaluated such an ontology for acute care to support interoperability, rule-based reasoning, and future decision support (40). Computational leak estimation can improve patient-flow reconstruction, a prerequisite for accurate waveform interpretation and automated cycling (15). Finally, reinforcement learning and digital twins may permit safety testing of ventilator policies before clinical exposure. Yu and colleagues proposed a safety-oriented offline reinforcement learning framework with temporal modeling and digital-twin verification for automated mechanical ventilation (41). Although not NIV-specific and currently a preprint, it illustrates why future algorithms must be uncertainty-aware, auditable, and validated under physiologically plausible scenarios.
For critical care NIV, AI-enabled closed-loop systems should be designed as supervised clinical tools rather than autonomous substitutes for clinicians. The clinician should remain in the loop, especially because NIV failure is a clinical diagnosis influenced by gas exchange, work of breathing, mental status, hemodynamics, secretion burden, airway protection, patient goals, and local expertise. AI may identify patterns that humans miss, but it must also be auditable, calibrated across populations, robust to missing data, and constrained by safety rules.
Discussion
Closed-loop NIV is best viewed as an emerging continuum rather than a mature therapeutic class. The most established components are leak-adapted triggering/cycling, dedicated NIV algorithms, and proportional or neural assistance. Direct NIV studies support improved synchrony and, in selected studies, comfort or fewer ventilator adjustments, but randomized NAVA/ASV evidence has not shown a consistent reduction in NIV failure, and the recent ASV-IntelliSync result remains feasibility evidence (4, 20–22, 29–33). Adaptive pressure modes such as AVAPS and iVAPS extend the concept by targeting ventilation, but outcome evidence remains uncertain and phenotype-dependent (38). In our view, the near-term role of closed-loop NIV should be bounded automation: machines can optimize rapidly changing technical tasks such as leak compensation, triggering, cycling, and proportional assistance, while clinicians retain control of therapeutic goals, escalation, and the decision to intubate. Table 1 summarizes these strategies, their feedback signals, automated outputs, clinical evidence, and main limitations.
This distinction is clinically important because NIV failure is not a single control error. Overreliance on automation may mask deterioration or delay intubation, whereas well-designed automation may reduce workload, practice heterogeneity, and delayed adjustment of settings. The invasive ventilation literature illustrates this balance: closed-loop systems can improve process quality without consistently improving patient-centered outcomes (18, 19). We therefore do not consider current evidence sufficient for an autonomous algorithm to determine continuation of NIV. Future studies should instead evaluate defined closed-loop functions in clinically coherent phenotypes, including hypercapnic COPD, obesity hypoventilation, cardiogenic pulmonary edema, post-extubation and postoperative support, and carefully selected hypoxemic respiratory failure.
Clinically relevant endpoints should include comfort, dyspnea, inspiratory effort relief, asynchrony burden, delivered tidal volume, gas-exchange stability, staff workload, intubation, delayed intubation, and patient-centered outcomes. A particularly informative trial would compare standard NIV with a supervised hierarchical strategy combining leak compensation, automated cycling, effort monitoring, and explicit failure-alert criteria. The system should disclose which feedback signal triggered an adjustment or alert and operate within predefined safety limits, allowing the clinician to distinguish technical optimization from a recommendation for escalation.
In summary, current evidence supports the physiological rationale for closed-loop NIV but not autonomous management or a major clinical-outcome benefit. We propose a hierarchical, multirate architecture in which fast technical loops manage leak and synchrony, intermediate loops regulate ventilation and unloading, and a slower supervisory layer integrates gas exchange, comfort, clinical trajectory, and failure risk. Such systems should remain clinician-supervised, transparent, and safety-constrained until prospective trials demonstrate that automation improves outcomes without delaying escalation.
Generative AI statement
The author(s) declared that generative AI was used in the creation of this manuscript. During the preparation of this manuscript, FG used OpenAI ChatGPT, GPT-5.5 Thinking model, OpenAI, to assist with literature organization, drafting support, language editing, reference checking, and preparation of the conceptual figure. OpenAI image-generation tools available within ChatGPT were also used to assist in creating Figure 1. The authors reviewed, edited, and verified all AI-assisted text, citations, references, and visual content, ensured that the figure accurately reflects the concepts discussed in the manuscript, and take full responsibility for the final content of the article. No generative AI tool is listed as an author.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Footnotes
Edited by: Marco Fiore, Università degli Studi della Campania 'Luigi Vanvitelli', Italy
Reviewed by: Ivaylo Minev, Plovdiv Medical University, Bulgaria
Statements
- Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The author AE declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.
- Funding
The author(s) declared that financial support was not received for this work and/or its publication.
- Author contributions
FG: Writing – review & editing, Conceptualization, Investigation, Writing – original draft, Project administration, Methodology. AE: Writing – review & editing, Project administration, Investigation, Writing – original draft, Methodology, Conceptualization.
References
- 1. Rochwerg B, Brochard L, Elliott MW, Hess D, Hill NS, Nava S. Official ERS/ATS clinical practice guidelines: noninvasive ventilation for acute respiratory failure. Eur Respir J. (2017) 50: 1602426. doi:10.1183/13993003.02426-2016
- 2. Rezoagli E, Nova A, Carteaux G, Giani M, Grieco DL, Pettenuzzo T. A clinical guide to non-invasive respiratory support in acute respiratory failure: ventilation settings, technical optimization and clinical indications. Crit Care. (2025) 29: 496. doi:10.1186/s13054-025-05730-y
- 3. Vignaux L, Tassaux D, Jolliet P. Performance of noninvasive ventilation modes on ICU ventilators during pressure support: a bench model study. Intensive Care Med. (2007) 33: 1444 – 51. doi:10.1007/s00134-007-0713-0
- 4. Carteaux G, Lyazidi A, Cordoba-Izquierdo A, Vignaux L, Jolliet P, Thille AW. Patient-ventilator asynchrony during noninvasive ventilation: a bench and clinical study. Chest. (2012) 142: 367 – 76. doi:10.1378/chest.11-2279
- 5. Caillard C, Fresnel E, Artaud-Macari E, Cuvelier A, Tamion F, Patout M. Ventilator performances for non-invasive ventilation: a bench study. BMJ Open Respir Res. (2024) 11: e002144. doi:10.1136/bmjresp-2023-002144
- 6. Perez J, Brandan L, Telias I. Monitoring patients with acute respiratory failure during non-invasive respiratory support to minimize harm and identify treatment failure. Crit Care. (2025) 29: 147. doi:10.1186/s13054-025-05369-9
- 7. Tonelli R, Fantini R, Tabbì L, Castaniere I, Pisani L, Pellegrino MR. Early inspiratory effort assessment by esophageal manometry predicts noninvasive ventilation outcome in de novo respiratory failure: a pilot study. Am J Respir Crit Care Med. (2020) 202: 558 – 67. doi:10.1164/rccm.201912-2512OC
- 8. Mocellin A, Guidotti F, Rizzato S, Tacconi M, Bruzzi G, Messina J. Monitoring and modulation of respiratory drive in patients with acute hypoxemic respiratory failure in spontaneous breathing. Intern Emerg Med. (2024) 19: 2105 – 19. doi:10.1007/s11739-024-03715-3
- 9. Von Platen P, Pomprapa A, Lachmann B, Leonhardt S. The dawn of physiological closed-loop ventilation-a review. Crit Care. (2020) 24: 121. doi:10.1186/s13054-020-2810-1
- 10. Chatburn RL, Mireles-Cabodevila E. Closed-loop control of mechanical ventilation: description and classification of targeting schemes. Respir Care. (2011) 56: 85 – 102. doi:10.4187/respcare.00967
- 11. Lellouche F, Brochard L. Advanced closed loops during mechanical ventilation (PAV, NAVA, ASV, SmartCare). Best Pract Res Clin Anaesthesiol. (2009) 23: 81 – 93. doi:10.1016/j.bpa.2008.08.001
- 12. Van Der Staay M, Chatburn RL. Advanced modes of mechanical ventilation and optimal targeting schemes. Intensive Care Med Exp. (2018) 6: 30. doi:10.1186/s40635-018-0195-0
- 13. Zheng M. Respiratory mechanics: revisiting the appraisement of lung recruitment. Respir Care. (2023) 68: 1262 – 70. doi:10.4187/respcare.10601
- 14. Parvinian B, Scully C, Wiyor H, Kumar A, Weininger S. Regulatory considerations for physiological closed-loop controlled medical devices used for automated critical care: food and drug administration workshop discussion topics. Anesth Analg. (2018) 126: 1916 – 25. doi:10.1213/ANE.0000000000002329
- 15. Vicario F, Alkhairy S, Buizza R, Truschel WA. Two-parameter leak estimation in non-invasive ventilation. In: Woo EJ, Zhang Y-t, Thomas Penzel T, editors. 2017 39th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC). Piscataway, NJ: Institute of Electrical and Electronics Engineers (IEEE) (2017). p. 153 – 6. doi:10.1109/EMBC.2017.8036785
- 16. Telias I, Spadaro S. Techniques to monitor respiratory drive and inspiratory effort. Curr Opin Crit Care. (2020) 26: 3 – 10. doi:10.1097/MCC.0000000000000680
- 17. Cammarota G, Simonte R, Longhini F, Spadaro S, Vetrugno L, De Robertis E. Advanced point-of-care bedside monitoring for acute respiratory failure. Anesthesiology. (2023) 138: 317 – 34. doi:10.1097/ALN.0000000000004480
- 18. Goossen RL, Schultz MJ, Tschernko E, Chew MS, Robba C, Paulus F. Effects of closed loop ventilation on ventilator settings, patient outcomes and ICU staff workloads: a systematic review. Eur J Anaesthesiol. (2024) 41: 438 – 46. doi:10.1097/EJA.0000000000001972
- 19. Sinnige JS, Buiteman-Kruizinga LA, Horn J, Paulus F, Schultz MJ, Serpa Neto A. Effect of automated closed-loop ventilation vs protocolized conventional ventilation on ventilator-free days in critically ill adults: a randomized clinical trial. JAMA. (2026) 335: 874 – 84. doi:10.1001/jama.2025.24384
- 20. Vignaux L, Vargas F, Roeseler J, Tassaux D, Thille AW, Kossowsky MP. Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study. Intensive Care Med. (2009) 35: 840 – 6. doi:10.1007/s00134-009-1416-5
- 21. Vignaux L, Tassaux D, Carteaux G, Roeseler J, Piquilloud L, Brochard L. Performance of noninvasive ventilation algorithms on ICU ventilators during pressure support: a clinical study. Intensive Care Med. (2010) 36: 2053 – 9. doi:10.1007/s00134-010-1994-2
- 22. Renaud Y, Auroi J, Cabrio D, Lupieri E, Chiche JD, Piquilloud L. Patient-ventilator synchrony under non-invasive ventilation is improved by an automated real-time waveform analysis algorithm: a bench study. Intensive Care Med Exp. (2025) 13: 16. doi:10.1186/s40635-025-00726-y
- 23. Hess DR. Patient-ventilator interaction during noninvasive ventilation. Respir Care. (2026) 71: 630 – 41. doi:10.1177/19433654251391165
- 24. Mojoli F, Pozzi M, Orlando A, Bianchi IM, Arisi E, Iotti GA. Timing of inspiratory muscle activity detected from airway pressure and flow during pressure support ventilation: the waveform method. Crit Care. (2022) 26: 32. doi:10.1186/s13054-022-03895-4
- 25. Mojoli F, Orlando A, Bianchi IM, Puce R, Arisi E, Salve G. Waveforms-guided cycling-off during pressure support ventilation improves both inspiratory and expiratory patient-ventilator synchronisation. Anaesth Crit Care Pain Med. (2022) 41: 101153. doi:10.1016/j.accpm.2022.101153
- 26. Calderini E, Confalonieri M, Puccio PG, Francavilla N, Stella L, Gregoretti C. Patient-ventilator asynchrony during noninvasive ventilation: the role of expiratory trigger. Intensive Care Med. (1999) 25: 662 – 7. doi:10.1007/s001340050927
- 27. Tassaux D, Gainnier M, Battisti A, Jolliet P. Impact of expiratory trigger setting on delayed cycling and inspiratory muscle workload. Am J Respir Crit Care Med. (2005) 172: 1283 – 9. doi:10.1164/rccm.200407-880OC
- 28. Chen Y, Cheng K, Zhou X. Effectiveness of inspiratory termination synchrony with automatic cycling during noninvasive pressure support ventilation. Med Sci Monit. (2016) 22: 1694 – 701. doi:10.12659/MSM.896059
- 29. Piquilloud L, Tassaux D, Bialais E, Lambermont B, Sottiaux T, Roeseler J. Neurally adjusted ventilatory assist improves patient-ventilator interaction during non-invasive ventilation delivered by face mask. Intensive Care Med. (2012) 38: 1624 – 31. doi:10.1007/s00134-012-2626-9
- 30. Longhini F, Pan C, Xie J, Cammarota G, Bruni A, Garofalo E. New setting of neurally adjusted ventilatory assist for noninvasive ventilation by facial mask: a physiologic study. Crit Care. (2017) 21: 170. doi:10.1186/s13054-017-1761-7
- 31. Prasad KT, Gandra RR, Dhooria S, Muthu V, Aggarwal AN, Agarwal R. Comparing noninvasive ventilation delivered using neurally-adjusted ventilatory assist or pressure support in acute respiratory failure. Respir Care. (2021) 66: 213 – 20. doi:10.4187/respcare.07952
- 32. Chhabria BA, Prasad KT, Dhooria S, Muthu V, Aggarwal AN, Agarwal R. A randomized controlled trial comparing non-invasive ventilation delivered using neurally adjusted ventilator assist (NAVA) or adaptive support ventilation (ASV) in patients with acute exacerbation of chronic obstructive pulmonary disease. J Crit Care. (2023) 75: 154250. doi:10.1016/j.jcrc.2022.154250
- 33. Vijay S, Sehgal IS, Agarwal R, Muthu V, Dhooria S, Prasad KT. Adaptive support ventilation with IntelliSync versus pressure support ventilation for non-invasive ventilation in acute exacerbation of COPD: a feasibility randomized trial. Respir Med. (2026) 253: 108690. doi:10.1016/j.rmed.2026.108690
- 34. Shah NM, D'Cruz RF, Murphy PB. Novel modes of non-invasive ventilation in chronic respiratory failure: a narrative review. J Thorac Dis. (2020) 12 (Suppl 2): S217 – 24. doi:10.21037/jtd-cus-2020-013
- 35. Briones Claudett KH, Esquinas Rodriguez A, Briones Claudett MH, Puga Tejada M, Cabrera Bańos MP, Daher N. J. Non-invasive mechanical ventilation with average volume-assured pressure support. Results according to the aetiology of acute respiratory failure. Anaesthesiol Intensive Ther. (2021) 53: 403 – 10. doi:10.5114/ait.2021.111527
- 36. Magdy DM, Metwally A. Effect of average volume-assured pressure support treatment on health-related quality of life in COPD patients with chronic hypercapnic respiratory failure: a randomized trial. Respir Res. (2020) 21: 64. doi:10.1186/s12931-020-1320-7
- 37. Magdy DM, Metwally A. Auto-titrating versus fixed-EPAP intelligent volume-assured pressure support (iVAPS) ventilation in patients with COPD and hypercapnic respiratory failure. Adv Respir Med. (2021) 89: 277 – 83. doi:10.5603/ARM.a2021.0056
- 38. Gregory B, Ezenwanne C, Al Mahrizi AD, Gill H, Mossolem F, Peng SA. AVAPS Versus BiPAP in acute hypercapnic respiratory failure: a systematic review and meta-analysis. COPD. (2026) 23: 2671706. doi:10.1080/15412555.2026.2671706
- 39. Carrasco-Ribelles LA, Llanes-Jurado J, Gallego-Moll C, Cabrera-Bean M, Monteagudo-Zaragoza M, Violán C. Prediction models using artificial intelligence and longitudinal data from electronic health records: a systematic methodological review. J Am Med Inform Assoc. (2023) 30: 2072 – 82. doi:10.1093/jamia/ocad168
- 40. Islam MF, Mosier J, Subbian V. Development and evaluation of an ontology for non-invasive respiratory support in acute care. PLoS One. (2026) 21: e0348199. doi:10.1371/journal.pone.0348199
- 41. Yu H, Liu H, Zhang Q, Joy W, Nikulina K, Schuppert AA. Ensuring safety in automated mechanical ventilation through offline reinforcement learning and digital twin verification. arXiv [Preprint]. (Accessed July 25, 2026) (2026). Available online at: https://arxiv.org/abs/2603.11372.
Citation
Guimaraes, F. S. & Esquinas, A. M. (2026). Closed-loop non-invasive ventilation in critical care: current evidence and clinical perspectives. Frontiers in Anesthesiology, Volume 5. https://doi.org/