Skip to content

Original Research

Frontiers in Oncology ·

Volume 16 - 2026 | https://doi.org/10.3389/fonc.2026.1960445

Total body irradiation practice patterns: insights from a national survey

  • Zaheeda Mulla1*corresponding author
  • Ali Balbaid1
  • Ghada Aldosary2,3
  • Wsam Ghandourh4
  • Ismail AlDahlawi5
  • Mohamed Bayoumy1
  • Mohannad Alhumaid2,3
  • Fahad Alrowais5
  • Hussain Almerdhemah1
  • Hiba Taha1
  • Abdullah AlMohamad1
  • Wasil Jastaniah1
  1. 1Department of Oncology, King Faisal Specialist Hospital and Research Centre, Jeddah, Saudi Arabia
  2. 2Radiation Oncology Section, Department of Oncology, King Abdullah Specialized Children's Hospital, Ministry of the National Guard-Health Affairs, Riyadh, Saudi Arabia
  3. 3King Abdullah International Medical Research Centre, Riyadh, Saudi Arabia
  4. 4Department of Health Administration and Medical Information, College of Health Science at Al-Leith, Umm Al-Qura University, Makkah, Saudi Arabia
  5. 5Department of Radiation Oncology, King Fahad Specialist Hospital-Dammam, Dammam, Saudi Arabia
Article metrics0 viewsView details

Abstract

Despite the expansion of hematopoietic stem cell transplantation (HSCT) services, contemporary Total Body Irradiation (TBI) practices remain insufficiently characterized in many healthcare settings. This study aimed to establish a national baseline for TBI practice to support benchmarking, standardization, and quality improvement. A cross-sectional online survey was distributed between May and June 2026 to seven Saudi centers that provided TBI services. The questionnaire collected data on service capacity, patient populations, dose fractionation, treatment techniques and delivery, planning methods, and quality assurance. A census sampling approach was used, inviting Radiation Oncologists, Medical Physicists, and Therapists. Data were analyzed using descriptive statistics in Python (v3.11). All seven centers responded. One center (C) submitted two responses covering conventional extended source-to-surface distance (SSD) and volumetric-modulated arc therapy (VMAT) practices. Commissioned between 1980 and 2020, the TBI programs manage an estimated 386 annual cases (range: 12–120). Fractionated TBI (10–12 Gy, six twice-daily fractions) predominated. The extended SSD opposed-beam technique remained in use at six centers, with the umbilicus as the most common prescription point. At two centers, extended-SSD TBI was used alongside Intensity-modulated radiation therapy (IMRT)-TBI (VMAT and Tomotherapy. IMRT-TBI-based centers reported volumetric target-coverage objectives of 90% and explicit organ-at-risk (OAR) dose planning constraints, particularly for the lungs and kidneys. Treatment planning for IMRT-TBI involved treatment planning systems (TPS), whereas extended-SSD techniques used varied planning approaches, including in-house spreadsheets, pre-calculated templates, and, in some cases, TPS-assisted planning. In vivo dosimetry was performed at all centers. The absorbed dose-rate at the prescription point ranged from <10 to 30 cGy/min for conventional techniques. Three centers planned their initial implementation of IMRT-TBI within three years. The ability to control OARs was the main driver, while linear accelerator availability and staff training were key barriers. Although conventional extended SSD TBI remains the most widely used technique, TBI practice is transitioning toward IMRT-based delivery. National practice guidelines based on existing international recommendations may help standardize care, address implementation barriers, and support the delivery of high-quality TBI services across treatment centers.

Keywords

Introduction

Total body irradiation (TBI), a specialized radiotherapy (RT) procedure, is a critical component of conditioning regimens for hematopoietic stem cell transplantation (HSCT) (1). Despite international guidelines advocating fractionated regimens and specifying dose and organ-at-risk (OAR) constraints to mitigate toxicity, substantial global variability persists in TBI delivery techniques, dose fractionation, shielding, and quality assurance (QA) protocols (2, 3).

Recent surveys have documented diverse institutional practices and highlighted an emerging technical shift from conventional lateral-opposed beam techniques to intensity-modulated radiation therapy (IMRT)-TBI (4–7). In this study, IMRT-TBI refers to TBI delivery using both volumetric modulated arc therapy (VMAT) and helical Tomotherapy-based treatment). The advantages of modern IMRT-based TBI techniques include improved sparing of several organs at risk, enhanced dose homogeneity, and more precise target dose delivery (8–11), however they remain underutilized due to resource availability, workflow challenges, limited clinical experience and lack of standardized protocols (7). Nonetheless, the technical and clinical advantages of VMAT-TBI, suggest that broader adoption is likely in the coming decade, making it essential for national surveys to track the pace and pattern of this transition.

Saudi Arabia’s expanding HSCT program lacks a systematic assessment of current TBI practice patterns, representing a significant knowledge gap and limiting opportunities for quality improvement, resource optimization, and adoption of modern techniques. International studies have demonstrated that national TBI practice surveys provide valuable insights into current practices, identify areas of consensus and variation, and establish benchmarks for quality improvement (4–7).

Methods and materials

Study design and population

A cross-sectional survey was conducted in May–June 2026, targeting all seven centers providing TBI services in Saudi Arabia, which were identified based on a recent study (12). Eligible participants included Radiation Oncologists, Medical Physicists, and Radiation Therapists, with a census sampling approach ensuring comprehensive multidisciplinary input.

Survey description and data collection

The survey, adapted from Li et al., assessed institutional capabilities, patient demographics, clinical indications, dose fractionations, treatment techniques, planning methods, QA procedures, and barriers to advanced technique adoption (4). Multiple-choice questions replaced open-ended items to improve data uniformity. The adapted survey was reviewed by a team of expert professionals representing each multidisciplinary team including a radiation oncologist, a physicist and a radiation therapist, before distribution.

Ethical considerations

The study was approved by the King Faisal Specialist Hospital and Research Centre Institutional Review Board (IRB#2261192), Saudi Arabia. Consent was obtained electronically from one designated respondent on behalf of each center prior to survey participation, and data confidentiality was maintained. No identifiers were used in this study as the survey was anonymous.

Data analysis

The data were cleaned and descriptively analyzed using Python (v3.11).

Results

TBI services background and general practice

All seven centers responded (100%) with a single multidisciplinary consensus response completed jointly by a radiation oncologist, medical physicist, and radiation therapist. Center C, however, submitted separate data for conventional lateral extended SSD and VMAT-TBI techniques to reflect the expanded clinical use of their IMRT-TBI. TBI services were commissioned between 1980 and 2020 (Figure 1). The reported experience with the technique varied: three centers had used it for 0–5 years, two for 5–10 years, and two for 15–20 years. IMRT-TBI started in 2025 in three centers (C, E and G). All centers confirmed VMAT capability of their linear accelerators with routine use in external beam radiotherapy treatment and with interest in developing national TBI guidelines.

Timeline of TBI commissioning in Saudi Arabia.
Figure 1. Timeline of TBI commissioning in Saudi Arabia.

Patient volume

Typical annual caseload estimates ranged from 12 to 120 unique patients, totaling 386 nationally, treated with either conventional technique or IMRT-TBI. Adult low-dose TBI (151patients) was the largest category. Center F reported the highest overall volume (n=120) and Center D reported the highest high-dose caseload (n=60 adults) (Figure 2). The definition of pediatric age varied from 0–12 to 0–18 years, most commonly 0–14 years (four centers).

Annual TBI patient volume by centers. Pediatric age cut-offs vary by center: ≤14y (A, D, E, F), ≤12y **(B)**, ≤16y **(C)**, ≤18y **(G)**.
Figure 2. Annual TBI patient volume by centers. Pediatric age cut-offs vary by center: ≤14y (A, D, E, F), ≤12y (B), ≤16y (C), ≤18y (G).

Clinical indications, dose and fractionation

Acute lymphoblastic leukemia (ALL) and acute myeloid leukemia (AML) were universal indications. The other indications are listed in Table 1. High-dose TBI (10–12 Gy in six twice-daily fractions) was the most commonly reported regimen for adults and pediatrics. Single-fraction low-dose regimens were also used.

Table 1. Clinical Indications and dose prescriptions for TBI at each center.

Clinical indicationsDose prescriptions
CenterAcute leukemia (ALL, AML)Sickle cell anemia & thalassemiaMyelodysplastic syndrome & multiple myelomaAutoimmune diseasesRefractory lymphomaAdultPediatric
A✓✓✗✓✗4Gy/1F2-3Gy/1F10-12Gy/6F BID4Gy/1F2-3Gy/1F
B✓✓✓✓✗10-12Gy/6F BID2-3Gy/1F10-12Gy/6F BID2–3 Gy/1F
C✓✗✗✗✓10-12Gy/6F BID4Gy/2F BID2–3 Gy/1F10-12Gy/6F BID
D✓✓✓✗✗10-12Gy/6F BID4Gy/2F BID4Gy/1F2-3Gy/1F10-12Gy/6F BID4Gy/2F BID4Gy/1F2-3Gy/1F
E✓✓✗✗✗10-12Gy/6F BID10-12Gy/6F BID
F✓✓✓✗✗10-12Gy/6F BID4Gy/2F BID4Gy/1F2-3Gy/1F10-12Gy/6F BID4Gy/2F BID4Gy/1F2–3 Gy/1F
G✓✗✗✗✗10-12Gy/6F BID10-12Gy/6F BID

BID, Twice daily dose; ALL, Acute lymphoblastic leukemia; AML, Acute myeloid leukemia.

For twice-daily fractionation, all centers used a minimum of 6 hours interfraction interval. The dose was most commonly prescribed to the umbilicus, either alone or in combination with another point for conventional TBI. For centers also using IMRT-TBI, a volume-based metric was utilized, covering 90% of the planned target volume (PTV)-body volume.

Dosimetry, planning and quality assurance

CT simulation was used in IMRT-TBI and non-CT approaches were used predominantly in conventional TBI (Table 2). When CT was used, body contouring was performed by medical physicists (typically <15 minutes), while target volume and OAR contouring was done by physicists, radiation oncologists, or both, with times ranging from under 15 to over 90 minutes. The most commonly contoured OARs were the lungs and lenses, followed by kidneys, heart, testicles, and ovaries (Table 2). The dose constraint values reported in Table 2 represent center-reported protocol parameters rather than treatment plan dosimetry derived from treated patients.

Table 2. Planning and dose constraints.

CenterCT-Sim usedTarget volumePlanning aims%OARs contouredLung dose constraint GyKidney dose constraint GyLensdose constraint GyOrgans shieldedShielding methodPlanning methodPlanning days
ANoBody+/-10N/AN/AN/AN/AN/AN/AIn-house spreadsheet2-3
BNoBody+/-10N/AN/AN/AN/ALungs, Heart, TesticlesLead BlocksIn-house spreadsheet3-6
C ExtendedSSDC VMATNoBody+/-10N/AN/AN/AN/AN/AN/APre-calculated template1-2
YesBody minus lung (cropped 3mm from skin)+/-15LungsKidneysLensMean Dose:<8Mean dose:≤10Max Dose:<12Lungs, Kidneys, Testicles, Lenses, BrainMLCsTPS> 6
DYesBody minus Lungs+/-10LungsMean Dose:<8N/AN/ALungsTesticlesLead BlocksTPS, In-house spreadsheet3-6
E ExtendedSSDE TomotherapyNoBody+/-10N/AN/AN/AN/ALungsRubber leadIn-house spreadsheet2-3
YesBody minus lungs, kidney, brain, and liver (Tomotherapy)+/-10LungsKidneysTesticles, Ovaries, Heart, LensMean Dose:<8Mean Dose: ≤10Max Dose:<12LungsKidneysMLCsTPS, In-house spreadsheet3-6
FYesBody+/-10LensBrain OAR structures for boost onlyMean Dose:<8N/AN/ALungsTesticlesLead BlocksTesticular shield (Pb)In-house spreadsheet2-3
GYesBody minus Lungs+/-5Lungs, Kidneys, Testicles, Ovaries, Heart, LensMean Dose:<8Mean Dose:<8Max Dose:<8LungsMLCsTPS1-2

VMAT, volumetric-modulated arc therapy; SSD, Source-to-Surface distance; N/A, Not applicable; MLCs, Multi-leaf collimators; TPS, Treatment planning system; CT-Sim, Computed tomography simulation; OARs, Organs at risk.

Treatment planning utilized treatment planning systems (TPS), in-house spreadsheets, or pre-calculated templates, taking 1 to over 6 days, and was performed exclusively by medical physicists. The reported planning objectives for dose homogeneity were ±10% (for all conventional techniques and Tomotherapy), ± 5% and ±15% for G and C, respectively, with VMAT-TBI. Refer to Table 2 for reported lung, kidney and lens dose constraints. Independent Excel-based dose checks were performed for all patients across four centers.

In vivo dosimetry was performed at all centers, most of which used 3–8 detectors. Detectors were placed at multiple sites, most consistently the head, lung, umbilicus, and pelvis, with additional sites (neck, mediastinum, thighs, knees, ankles and under the arms) varying by center.

Treatment delivery technique and setup

The most prevalent delivery technique reported was extended SSD using lateral opposed beams. Two centers employed VMAT. Center E used lateral beams, a moving couch technique, and Tomotherapy (for patient length < 135 cm). The other delivery techniques and energies used are listed in Table 3. The extended SSD ranged from 286 to 455 cm (212 cm for the moving couch), while VMAT and Tomotherapy centers treated at SAD of 100 and 85 cm, respectively. The supine position, was the most commonly used treatment position.

Table 3. Treatment delivery parameters.

CenterDelivery techniquePhoton Beam energyMVPatient positionBuild-up materialsSSD or SAD setupSAD/SSD distancecmPre-treatment imagingTypical dose ratecGy/minDose rate Measured/Calculated
ALateral beams at extended SSD15Supine,Supine with knees bentRice BagsPerspex Beam SpoilerBolusSSD286None<10Calculated
BAP beams at extended SSD6Lying lateralPerspex Beam SpoilerSSD400None<10Both
CLateral beams at extended SSD18Supine,Supine with knees bentRice BagsPerspex Beam SpoilerSSD455None10-20Both
VMAT6Supine,Supine with knees bentN/ASAD100Linac-based 2D/3D IGRT (i.e., EPID, kV, CBCT)20-30Calculated
DAP beams at extended SSDLateral beams at extended SSD15Supine,Lying lateralRice BagsPerspex Beam SpoilerBolusSSD370Portable X-rays (detector or unit)10-20Calculated
ELateral beams at extended SSD6Supine with knees bent, SupineRice BagsPerspex Beam SpoilerSSD400Linac-based 2D IGRT (i.e., EPID)<10Calculated
Moving couch technique6Supine with knees bent, SupineRice BagsPerspex Beam SpoilerSSD212Linac-based 2D20-30Calculated
Centralize Tomotherapy6 FFFSupineN/ASAD85MV-CBCT<87Calculated
FLateral beams at extended SSD10SupineRice BagsPerspex Beam SpoilerSSD450None<10Calculated
GLinac-based VMAT6SupineRice BagsSAD100Linac-based 2D IGRT (i.e., EPID)Linac-based 3D IGRT (i.e., CBCT)<10Measured

SSD, Source-to-Surface distance; SAD, Source-to-Axis distance; MV, Megavoltage; MU, Monitor unit; Linac, Linear accelerator; CBCT, Cone-beam computed tomography; VMAT, Volumetric-modulated arc therapy; AP, Anterior-posterior; IGRT, Image-guided radiation therapy; EPID, Electronic portal imaging device.

Lung shielding was the most widely applied OAR protection reported; In two extended-SSD approaches, no physical lung shielding was used; instead, lung dose was modified through arm positioning and bolus application. Testicular shielding was reported for three approaches, kidney shielding for two, and heart, lens, and brain shielding for one approach each, with the latter reported for VMAT (Table 2). The most common method utilized for shielding was lead blocks, followed by multileaf collimator (MLC) shielding (VMAT technique) and lead rubber. Patient positioning included custom or in-house TBI beds, stretchers with custom mattresses, linac couch tops for VMAT, and aids such as head supports, Vacloc bags, and masks.

The reported absorbed dose rates at the prescription point for conventional and VMAT-TBI ranged from <10 cGy/min to 30 cGy/min, with a dose rate of <87 cGy/min reported for Tomotherapy at Center E. The dose rate values reported by the centers were determined by calculation, both measurement and calculation, or measurement alone (Table 3).

In contrast, the reported linac dose-rate settings ranged from 100 to 400 MU/min for conventional techniques, while Center C reported a reduced setting of 200 MU/min for chest arcs and 600 MU/min for other body regions during VMAT-TBI. Adult delivery times varied, with the longest reported for VMAT at Center C and generally shorter delivery times reported for pediatric patients.,

Pre-treatment imaging included portable X-rays for conventional TBI and IGRT for IMRT-TBI. (Table 3).

Self-reported practice ratings

Ten aspects of practice were self-rated from 1 (poor) to 5 (excellent). Linac reliability and intra-departmental workflow scored the highest (mean 4.5 each), followed by dose delivery accuracy and in vivo dosimetry (4.2 each). OAR dose control showed the greatest variation (1–5, mean 3.5), reflecting the contrast between conventional centers with limited shielding and VMAT techniques with full optimization. The availability of backup-technique also varied (3–5, mean 3.9). Patient experience and dose uniformity accuracy each averaged 4.1, with scores as low as 3. (Figure 3).

Self-rated clinical practice parameters by technique profile. Seven institutions contributed data, with Center C providing separate responses for its conventional extended-SSD technique (C1) and VMAT-TBI technique (C2), resulting in eight technique-specific profiles. A, B, C1, D, and F represented conventional extended-SSD practice; C2, E and G represented IMRT-TBI.
Figure 3. Self-rated clinical practice parameters by technique profile. Seven institutions contributed data, with Center C providing separate responses for its conventional extended-SSD technique (C1) and VMAT-TBI technique (C2), resulting in eight technique-specific profiles. A, B, C1, D, and F represented conventional extended-SSD practice; C2, E and G represented IMRT-TBI.

Future readiness

Three of the seven centers planned their first implementation of IMRT-TBI within three years, except Center F. Center C was already using VMAT as its increasingly established default technique, while Centers E and G were using IMRT-TBI for selected cases. The most cited improvement priority was OAR dose control, followed by patient experience and prescription dose uniformity (Figure 4). The main barriers to VMAT adoption were linac slot availability and training (three centers each), interdepartmental workflow (two centers), and staffing and patient volume constraints (two centers each). (Figure 5).

Aspects being actively improved.
Figure 4. Aspects being actively improved.
Perceived barriers to VMAT adoption for TBI.
Figure 5. Perceived barriers to VMAT adoption for TBI.

Discussion

Our survey provides the first national overview of TBI in Saudi Arabia. While several commonalities were identified, the participating centers provided a valuable opportunity to examine variations and evolving trends in TBI practice across Saudi Arabia.

Patient volume and service growth

The combined annual TBI caseload across the participating centers highlights the established role of TBI within the national radiotherapy landscape. Variability in institutional experience is consistent with international practice, where TBI services are often delivered in specialized transplant centers (4). The data demonstrated that adult low-dose TBI was the most common treatment category, followed by adult high-dose, pediatric high-dose, and pediatric low-dose TBI, reflecting the diverse clinical indications for TBI.

The establishment timeline of TBI services (spanning from 1980 to 2020) illustrates progressive national capacity building, with three centers reporting the commissioning of IMRT-TBI in 2025. While pediatric age definitions have historically varied, the Saudi Health Council recently mandated a standardized definition for individuals under 18 years of age (13).

Clinical indications and dose fractionation

Patients with ALL and AML were treated across all Saudi centers, consistent with the well-established role of TBI as a conditioning regimen for HSCT in patients with hematologic malignancies. The breadth of indications, included hemoglobinopathies and autoimmune diseases, reflecting both the diversity of transplant protocols and regional disease epidemiology, particularly the high prevalence of hemoglobinopathies in Saudi Arabia and the surrounding region (14, 15).

High-dose fractionated TBI (10–12 Gy in six fractions delivered twice daily) adult regimen was used by all but one center. However, all centers applied this regimen to pediatric patients. The Canadian survey reported near-identical fractionation dominance, with 12 Gy in 6 fractions used by 11 of the 12 TBI-performing centers. Notably, half of the Canadian centers also reported using a dose prescription unique to their facility, reflecting a similar degree of institutional variation to that observed in Saudi Arabia (Table 1; Figure 3) (5).

The six-hour minimum inter-fraction interval reported by all centers is consistent with radiobiological principles, allowing adequate repair of sublethal damage in normal tissues between fractions (2, 16). Low-dose single-fraction TBI (2–4 Gy) was used by most centers for adults and pediatrics, typically for a reduced-intensity conditioning regimen. This fractionation heterogeneity is consistent with international practices, where regimen selection is dictated by the transplant protocol, disease type, patient age, and comorbidities (3, 17). Unlike malignant indications, which commonly employ fractionated TBI to a total dose of 12 Gy, conditioning regimens for thalassemia and aplastic anemia frequently incorporate low dose TBI (18).

The diversity in dose prescription points and volume-based PTV-Body D90% reflects the transition from conventional point-based dosimetry to modern volumetric IMRT planning, where dose prescription to a volumetric metric enables a more rigorous evaluation of target coverage and homogeneity compared to traditional point-dose specifications (1).

Technique variation and the transition to VMAT

The dominance of conventional extended SSD lateral beam techniques reflects the historically established clinical practice, while the emerging VMAT adoption signals modernization aligned with global trends (2). A systematic comparison with current international guidelines contextualizes this finding. The ACR–ARS practice parameter for the performance of TBI addresses both conventional and evolving volumetric modulated TBI approaches, reflecting the continued role of conventional techniques alongside the development of conformal delivery methods (1). The International Lymphoma Radiation Oncology Group [ILROG] consensus guidelines support the use of CT-based volumetric planning and conformal techniques for TBI, with the aim of improving dose homogeneity and facilitating organ-at-risk sparing (2). The European Society for Radiotherapy and Oncology (ESTRO) ACROP and European Society for Paediatric Oncology (SIOPE) recommendations, particularly for pediatric patients, similarly support highly conformal techniques such as IMRT-TBI to improve dose homogeneity, reduce radiation exposure to organs at risk, and thereby mitigate potential late effects in developing organs (3).

Among the three centers utilizing IMRT-based TBI, Center C reported an expanded adoption of VMAT-TBI in routine clinical practice whereas Centers E and G are using IMRT-TBI for selected patients and are currently in a transition phase. The Saudi baseline stands in contrast to the 2024 ANZ survey in which seven of 19 centers had already adopted modulated techniques (VMAT or helical Tomotherapy), while 12 continued to use large open fields (4). Regionally, the broader regional landscape mirrors this shift toward modernization; the King Hamad University Hospital Oncology Center in Bahrain and the Sultan Qaboos Comprehensive Cancer Care and Research Center (SQCCCRC) in Oman have introduced VMAT-TBI, while the Kuwait Cancer Control Center utilizes extended SSD modulated TBI (19).

Lung shielding is the most commonly used organ protection measure. Notably, the shielding of the lungs, kidneys, testicles, lenses and brain in Center C with VMAT application reflects the organ-sparing capabilities of VMAT-based TBI techniques (20).

Dosimetric studies comparing VMAT, helical Tomotherapy, and conventional techniques have demonstrated that conformal approaches reduce the mean lung dose by approximately 25% compared to 2D AP/PA techniques, with VMAT and Tomotherapy achieving lung mean doses in the 6–8 Gy range versus higher values with conventional lateral beams (10, 21, 22). VMAT has also been shown to increase PTV D90% by 6–7% compared with 2D techniques and provide additional sparing of the kidneys, gonads, brain, and thyroid (10, 21). However, with a lack of prospective data comparing conventional vs IMRT-TBI, it is not clear whether dosimetric advantages will translate into better outcomes.

Dosimetry, planning, and quality assurance

CT simulation was used for all IMRT-TBI techniques, while conventional techniques primarily used non-CT-based approaches. International guidance increasingly recognizes CT-based three-dimensional planning as an approach that can facilitate assessment of dose distribution, dose heterogeneity, and organ-at-risk exposure, particularly with conformal techniques. The ILROG guidelines describe the increasing use of CT-based 3-D planning and conformal techniques, while also recognizing conventional approaches based on point measurements (2). The most commonly contoured OARs were the lungs and lenses, followed by the kidneys, heart, testicles, ovaries. This pattern aligns with international consensus recommendations, which emphasize the lung mean dose as the primary constraint for high-dose TBI, with additional consideration of the kidneys, lenses, and gonads (1, 23). Even when active sparing is not performed, contouring enables dose documentation, protocol compliance, and future evaluation of dose–toxicity relationships.

Treatment planning utilized TPS, in-house spreadsheets, or templates, with planning times varying from 1 to over 6 days, (for VMAT-TBI in Center C), reflecting the substantial time investment required for conformal TBI planning. Published institutional experiences report contouring times of several hours and optimization/dose calculation times spanning many hours to days for VMAT-TBI, with QA procedures adding several additional hours per patient (24–26). This workflow burden represents a major barrier to the broader adoption of VMAT (4).

According to the American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) guidelines, dose inhomogeneity should generally be maintained within ±10% (1, 27), to ensure adequate target coverage while limiting excessive hot and cold spots. However, in VMAT-TBI, greater dose inhomogeneities may be acceptable than with conventional TBI to balance target coverage with organ-at-risk sparing (1). The ILROG similarly endorses ±10% homogeneity for conventional TBI but acknowledges that planned heterogeneity may be intentionally introduced in conformal techniques to achieve OAR sparing (2). The ESTRO/SIOPE recommendations permit selective dose reduction to OARs provided that adequate target coverage is maintained (3). Dose homogeneity of ±10% was the predominant planning objective among surveyed Saudi centers.

While the reported QA practices demonstrated strengths in the universal adoption of in vivo dosimetry to assess the dose delivered to the patient, independent dose verification, involving an additional check of the calculated dose using a separate calculation method or system, was performed in only half the centers, highlighting a key area for improvement. ILROG guidelines recommend multi-layered QA encompassing patient-specific pre-treatment verification and independent secondary dose calculation (2), and recent VMAT-TBI and Tomotherapy institutional series consistently implement portal dosimetry or diode/array measurements together with secondary calculation checks as standard (24, 25, 28, 29).

Organ-at-risk dose control and clinical implications

Self-reported practice ratings revealed that OAR dose control showed the greatest variation. The ACR-ARS practice parameter highlights the evaluation of the mean lung dose as an important factor related to pulmonary toxicity for high-dose TBI regimens (1), and clinical reports emphasize lung sparing as a primary driver for adopting conformal techniques (21–23).

Early clinical series of linac-based VMAT and Tomotherapy TBI reported acceptable acute and subacute toxicity, with one cohort demonstrating reduced severe mucositis after implementing oropharyngeal mucosa sparing (30). The clinical value of the organ sparing is most clearly established for the lungs, where reductions in mean lung dose have been associated with lower rates of radiation pneumonitis (2, 31). Consequently, a mean lung dose constraint of approximately 8 Gy has been widely adopted in contemporary TBI protocols and is readily achievable with modulated TBI techniques (1, 3). The centers employing IMRT-TBI techniques reported planning objectives of a mean lung of <8 Gy and mean renal dose of <8–10 Gy and maximum lens of <8–12 Gy for adults and pediatrics, consistent with ACR–ARS, ILROG, and ESTRO/SIOPE thresholds (1–3).

The centers that identified OAR dose control as an area for active improvement indicated the perceived limitations of OAR sparing provided by conventional techniques. This finding, combined with the fact that many centers plan to adopt IMRT-TBI within three years, suggests that Saudi TBI practice is set to transform in the near future. However, replacing extended SSD with VMAT-TBI increases resource demand. A key barrier is that limited couch travel restricts the maximum treatable length in the superior–inferior direction, requiring patient repositioning between head-first (HF) and feet-first (FF) supine orientations in centers without rotational tabletops, increasing planning and treatment delivery owing to multiple isocenters and extensive image guidance (10). Furthermore, the use of multiple abutting fields complicates field matching, often introducing dose uncertainties at the junction interfaces. Additionally, the shift away from simultaneous whole-body irradiation has prompted concerns regarding the potential under-dosage of circulating leukemia cells due to the temporal gap in delivery (2).

Self-rated practice and quality metrics

Dose delivery accuracy and in vivo dosimetry received consistently high self-rated scores, indicating a high perceived confidence level in dosimetric verification practices. Patient experience and dose uniformity accuracy also received high mean ratings. This was supported by qualitative feedback from Center B, which described extended SSD as “a reliable setup that ensures a positive experience for patients and staff, while maintaining dose delivery and accuracy.” However, both measures received low ratings from Center C, indicating variability in perceived performance across the centers. These findings represent the respondents’ perceptions of practice and should not be interpreted as evidence of objectively measured dose accuracy or uniformity.

The availability of backup techniques showed considerable variation, highlighting potential vulnerabilities in service resilience. Given the time-sensitive nature of transplant conditioning (2, 32), maintaining continuity of TBI services during equipment failure or unplanned downtime is essential (20). The development of contingency pathways, including backup treatment protocols (1) and inter-institutional referral agreements, may represent an important area for future national coordination.

Future readiness and barriers to VMAT adoption

VMAT adoption within three years was the most commonly cited target. Center G has begun VMAT-TBI for patients <150 cm and plans to commission a rotating tabletop for taller patients in the future. Of the centers not planning near-term changes, Center C had already implemented and expanded their VMAT-TBI program, and Center F indicated that adoption of a new technique was not planned, stating that “our extended SSD technique is generally robust and reliable”.

The perceived barriers most commonly reported in VMAT adoption; linac slot availability, training, interdepartmental workflow challenges, staffing, and patient volume constraints are consistent with international reports (4, 22, 24–26, 33) and represent systemic challenges in radiotherapy. A national survey confirms this, with nearly 60% of departments reporting limited technical skills training, underscoring the critical need for workforce development and workflow integration (12).

Proposed solutions include auto-segmentation, automated planning, standardized recommendations, QA frameworks, and multicenter efforts to align Dose-Volume Histogram and OAR constraint practices (20, 23, 24, 34).

The need for national guidelines

All Saudi centers supported the development of national TBI guidelines, reflecting the considerable variation in techniques, dosimetry, planning objectives, OAR constraints, QA practices, and dose prescription points observed across centers. Comparable international variations have similarly driven calls for standardization (5, 33). Similar international variation has prompted standardization efforts and published recommendations (3, 20, 23, 35). A national multidisciplinary TBI steering committee could adapt these guidelines to local practice while supporting workforce planning, benchmarking, quality improvement and equitable access to advanced TBI techniques.

Limitations and future directions

Reliance on self-reported data and the mixture of institution-level and technique-level observations represent important limitations of this study. Potential differences in the interpretation of questions, together with the inclusion of centers undergoing practice transition at the time of the survey, represent important considerations when interpreting the findings. Other caveats include the lack of independent verification of reported data, variation in pediatric age definitions and the absence of patient-level dosimetric or clinical outcome data. Nevertheless, these findings provide a national baseline for prospective multicenter studies and support the establishment of a national TBI registry for longitudinal quality monitoring, research, and international collaboration.

Conclusion

This survey of centers employing TBI practices in Saudi Arabia revealed a service landscape characterized by substantial patient volume, diverse clinical indications, and significant technical heterogeneity. The dominance of conventional lateral beam techniques at extended SSD, combined with widespread plans to adopt VMAT within three years at most centers, positions Saudi Arabia’s TBI practice at a critical inflection point. Building on existing international recommendations, a nationally coordinated approach could support the standardization of practice, to support the delivery of high-quality TBI services across treatment centers for patients undergoing HSCT.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fonc.2026.1960445/full#supplementary-material

Supplementary material Supplementary material at the source

Footnotes

Edited by: Thomas FitzGerald, University of Massachusetts Boston, United States

Reviewed by: Giuseppe Schettino, National Physical Laboratory, United Kingdom

Chenlei Guo, National Cancer Center/Chinese Academy of Medical Sciences/Peking Union Medical College Cancer Hospital Surgery, China

Sevim Sahin, Fenerbahçe University, Türkiye

Statements

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Funding

The author(s) declared that financial support was not received for this work and/or its publication.

Data availability

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics approval

The studies involving humans were approved by KFSH&RC Research Ethics Committee. The studies were conducted in accordance with the local legislation and institutional requirements. The ethics committee/institutional review board waived the requirement of written informed consent for participation from the participants or the participants’ legal guardians/next of kin because Consent was obtained electronically prior to survey participation, with data confidentiality maintained.

Author contributions

ZM: Writing – review & editing, Validation, Conceptualization, Methodology, Project administration, Writing – original draft, Visualization, Data curation. AB: Data curation, Writing – review & editing, Conceptualization, Validation, Methodology, Writing – original draft. GA: Writing – review & editing, Writing – original draft, Validation, Data curation, Methodology, Conceptualization. WG: Formal analysis, Validation, Methodology, Writing – review & editing, Writing – original draft. IA: Methodology, Data curation, Validation, Conceptualization, Writing – review & editing, Writing – original draft. MB: Writing – review & editing, Methodology, Conceptualization, Writing – original draft, Visualization, Validation. MA: Methodology, Writing – review & editing, Writing – original draft, Conceptualization, Validation. FA: Validation, Conceptualization, Methodology, Writing – review & editing, Writing – original draft, Visualization. HA: Writing – original draft, Data curation, Conceptualization, Methodology, Writing – review & editing. HT: Conceptualization, Writing – review & editing, Writing – original draft, Data curation, Methodology. AA: Conceptualization, Writing – original draft, Methodology, Writing – review & editing, Data curation. WJ: Visualization, Methodology, Conceptualization, Supervision, Validation, Writing – original draft, Writing – review & editing.

Acknowledgements

The authors would like to thank Sarah Li (sarah.li@health.qld.gov.au) for providing us with the survey used in their study.

References

  1. 1 Reilly M, Dandapani SV, Kumar KA, Constine L, Fogh SE, Roberts KB. ACR–ARS Practice Parameter for the Performance of Total Body Irradiation. Am J Clin Oncol. (2023) 46: 185–92. doi: doi:10.1097/COC.0000000000000997
  2. 2 Wong JYC, Filippi AR, Dabaja BS, Yahalom J, Specht L. Total body irradiation: Guidelines from the International Lymphoma Radiation Oncology Group (ILROG). Int J Radiat Oncol Biol Phys. (2018) 101: 521–9. doi: doi:10.1016/j.ijrobp.2018.04.071
  3. 3 Hoeben BAW, Pazos M, Seravalli E, Bosman ME, Losert C, Albert MH. ESTRO ACROP and SIOPE recommendations for myeloablative Total Body Irradiation in children. Radiotherapy Oncol. (2022) 173: 119–33. doi: doi:10.1016/j.radonc.2022.05.027
  4. 4 Li S, Hamad N, Le TT, Glassborow E, Kupfer T, Khor R. Total Body Irradiation in Australia and New Zealand: A 2024 Practice Survey on Changing Patterns of Care. J Med Imaging Radiat Oncol. (2026) 70 (3): 322–30. doi: doi:10.1111/1754-9485.70074
  5. 5 Studinski RCN, Fraser DJ, Samant RS, MacPherson MS. Current practice in total-body irradiation: Results of a Canada-wide survey. Curr Oncol. (2017) 24: 181–6. doi: doi:10.3747/co.24.3484
  6. 6 Giebel S, Miszczyk L, Slosarek K, Moukhtari L, Ciceri F, Esteve J. Extreme heterogeneity of myeloablative total body irradiation techniques in clinical practice: A survey of the Acute Leukemia Working Party of the European Group for Blood and Marrow Transplantation. Cancer. (2014) 120: 2760–5. doi: doi:10.1002/cncr.28768
  7. 7 Fog LS, Wirth A, MacManus M, Downes S, Grace M, Moggré A. Total body irradiation in Australia and New Zealand: results of a practice survey. Phys Eng Sci Med. (2020) 43: 825–35. doi: doi:10.1007/s13246-020-00878-z
  8. 8 Hoeben BAW, Pazos M, Albert MH, Seravalli E, Bosman ME, Losert C. Towards homogenization of total body irradiation practices in pediatric patients across SIOPE affiliated centers. A survey by the SIOPE radiation oncology working group. Radiotherapy Oncol. (2021) 155: 113–9. doi: doi:10.1016/j.radonc.2020.10.032
  9. 9 Sun R, Cuenca X, Itti R, Nguyen Quoc S, Vernant J-P, Mazeron J-J. First French experiences of total body irradiations using helical TomoTherapy®. Cancer/Radiotherapie. (2017) 21: 365–72. doi: doi:10.1016/j.canrad.2017.01.014
  10. 10 Kovalchuk N, Simiele E, Skinner L, Yang Y, Howell N, Lewis J. The Stanford Volumetric Modulated Arc Therapy Total Body Irradiation Technique. Pract Radiat Oncol. (2022) 12: 245–58. doi: doi:10.1016/j.prro.2021.12.007
  11. 11 Ahn KH, Rondelli D, Koshy M, Partouche JA, Hasan Y, Liu H. Knowledge-based planning for multi-isocenter VMAT total marrow irradiation. Front Oncol. (2022) 12: 942685. doi: doi:10.3389/fonc.2022.942685
  12. 12 Mulla Z, Ghandourh W, Almerdhemah H, Albish N, Kattan A, Qahtani AA. Unlocking radiotherapy access: the critical role of structural quality and radiation therapists in Saudi Arabia. Front Oncol. (2026) 16: 1802072. doi: doi:10.3389/fonc.2026.1802072
  13. 13 Saudi Health Council. Saudi Health Council Official Website. Available online at: https://shc.gov.sa/En (Accessed July 17, 2026).
  14. 14 Memish ZA, Owaidah TM, Saeedi MY. Marked regional variations in the prevalence of sickle cell disease and β-thalassemia in Saudi Arabia: Findings from the premarital screening and genetic counseling program. J Epidemiol Glob Health. (2011) 1: 61–8. doi: doi:10.1016/j.jegh.2011.06.002
  15. 15 Aljurf M, Nassar A, Hamidieh AA, Elhaddad A, Hamladji R-M, Bazarbachi A. Hematopoietic stem cell transplantation in the Eastern Mediterranean Region (EMRO) 2011-2012: A comprehensive report on behalf of the Eastern Mediterranean Blood and Marrow Transplantation group (EMBMT). Hematology/Oncology Stem Cell Ther. (2015) 8: 167–75. doi: doi:10.1016/j.hemonc.2015.09.002
  16. 16 Muller-Runkel R, Vijayakumar S. Fractionated total body irradiation: Some radiobiological considerations. Clin Oncol. (1993) 5: 39 – 42. doi: doi:10.1016/S0936-6555(05)80695-8
  17. 17 Keit E, Liveringhouse C, Figura N, Weygand J, Sandoval ML, Garcia G. Feasibility and Toxicity of Full-Body Volumetric Modulated Arc Therapy Technique for High-Dose Total Body Irradiation. Technol Cancer Res Treat. (2023) 22: 15330338231180779. doi: doi:10.1177/15330338231180779
  18. 18 DeZern AE, Zahurak ML, Symons HJ, Cooke KR, Rosner GL, Gladstone DE. Haploidentical BMT for severe aplastic anemia with intensive GVHD prophylaxis including posttransplant cyclophosphamide. Blood Adv. (2020) 4: 1770–79. doi: doi:10.1182/bloodadvances.2020001729
  19. 19 Kumar SAP, Hassan YM, Bushehri A, Fadel S, Khan MI, Narayanan D. Streamlining total body irradiation (TBI) patient setup: an innovative approach for homogenous dose distribution, enhanced reproducibility, and improved patient outcomes in HSCT settings using an extended technique. ESTRO 2026 Congress. Stockholm, Sweden (2026). p. Abstract No. 2004.
  20. 20 Seravalli E, Bosman ME, Han C, Losert C, Pazos M, Engström PE. Technical recommendations for implementation of Volumetric Modulated Arc Therapy and Helical Tomotherapy Total Body Irradiation. Radiotherapy Oncol. (2024) 197: 110366. doi: doi:10.1016/j.radonc.2024.110366
  21. 21 Ngo N, Blomain ES, Simiele E, Romero I, Hoppe RT, Hiniker SM. Improved organ sparing using auto‐planned Stanford volumetric modulated arc therapy for total body irradiation technique. Pediatr Blood Cancer. (2023) 70: e30589. doi: doi:10.1002/pbc.30589
  22. 22 Loginova AA, Tovmasian DA, Lisovskaya AO, Kobyzeva DA, Maschan MA, Chernyaev AP. Optimized Conformal Total Body Irradiation methods with Helical TomoTherapy and Elekta VMAT: Implementation, Imaging, Planning and Dose Delivery for Pediatric Patients. Front Oncol. (2022) 12: 785917. doi: doi:10.3389/fonc.2022.785917
  23. 23 Kovalchuk N, Simiele EA, LaRiviere M, Hiniker SM, Soike M, Han C. Incorporating Intensity Modulated Total Body Irradiation (IMRT-TBI) Into Future Cooperative Group Clinical Trials: An NRG Hematologic Malignancies Working Group-Led Report From the National Clinical Trials Network. International Journal of Radiation Oncology*Biology*Physics. (2026) 125: 714–36. doi: doi:10.1016/j.ijrobp.2026.02.201
  24. 24 Springer A, Hammer J, Winkler E, Track C, Huppert R, Böhm A. Total body irradiation with volumetric modulated arc therapy: Dosimetric data and first clinical experience. Radiat Oncol. (2016) 11: 46. doi: doi:10.1186/s13014-016-0625-7
  25. 25 La Macchia G, Wan C, Dass J, Taylor M, Neveri G, Skorska M. Technique considerations for implementing volumetric‐modulated arc therapy for total body irradiation within an Australian tertiary institution. J Med Radiat Sci. (2025) 72: 165–72. doi: doi:10.1002/jmrs.844
  26. 26 Haertter A, Goel K, Niculcea J, Li T, Hill-Kayser CE, Plastaras J. Dosimetric Performance Analysis of Pediatric Based HalcyonTM Volumetric Arc Therapy-Based Lung-Sparing Total Body Irradiation. Technol Cancer Res Treat. (2026) 25: 15330338261443068. doi: doi:10.1177/15330338261443068
  27. 27 Wolden SL, Rabinovitch RA, Bittner NHJ, Galvin JM, Giap HB, Schomberg PJ. American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) Practice Guideline for the Performance of Total Body Irradiation (TBI). American Journal of Clinical Oncology. (2013) 36: 97–101. doi: doi:10.1097/COC.0b013e31826e0528
  28. 28 Kakade NR, Kumar R, More M, Chakrabarty S, Sharma SD, Sapra BK. Dosimetric analysis of VMAT-TBI: Measurement of organ and junction dose using a whole body anthropomorphic phantom. Radiat Environ Biophys. (2026). doi: doi:10.1007/s00411-026-01213-y
  29. 29 Morrison CT, Symons KL, Woodings SJ, House MJ. Verification of junction dose between VMAT arcs of total body irradiation using a Sun Nuclear Arc CHECK phantom. J Appl Clin Med Phys. (2017) 18: 177–82. doi: doi:10.1002/acm2.12208
  30. 30 Köksal M, Özkan O, Holderried T, Heine A, Brossart P, Gawish A. Optimized Conformal Total Body Irradiation with VMAT Using a Linear-Accelerator-Based Radiosurgery Treatment System in Comparison to the Golden Standard Helical TomoTherapy. Cancers. (2023) 15: 4220. doi: doi:10.3390/cancers15174220
  31. 31 Hui C, Simiele E, Lozko Y, Romero I, Skinner L, Binkley MS. Volumetric modulated arc therapy total body irradiation improves toxicity outcomes compared to 2D total body irradiation. Front Oncol. (2024) 14: 1459287. doi: doi:10.3389/fonc.2024.1459287
  32. 32 Gyurkocza B, Sandmaier BM. Conditioning regimens for hematopoietic cell transplantation: One size does not fit all. Blood. (2014) 124: 344–53. doi: doi:10.1182/blood-2014-02-514778
  33. 33 Misson-Yates S, Cunningham R, Gonzalez R, Diez P, Clark CH. Optimised conformal total body irradiation: A heterogeneous practice, so where next ? BJR. (2023) 96: 20220650. doi: doi:10.1259/bjr.20220650
  34. 34 Frederick R, Van Dyke L, Hudson A, Pierce G. Advanced automated treatment planning for total body irradiation: Implementation and effects on standardization. Physica Med. (2023) 112: 102623. doi: doi:10.1016/j.ejmp.2023.102623
  35. 35 Hoeben BAW, Saldi S, Aristei C, Engellau J, Ocanto A, Hiniker SM. Rationale, implementation considerations, delineation and planning target objective recommendations for volumetric modulated arc therapy and helical tomotherapy total body irradiation, total marrow irradiation, total marrow and lymphoid irradiation and total lymphoid irradiation. Radiotherapy and Oncology. (2025) 206: 110822. doi: doi:10.1016/j.radonc.2025.110822

Citation

Mulla, Z., et al. (2026). Total body irradiation practice patterns: insights from a national survey. Frontiers in Oncology, Volume 16. https://doi.org/
Received
Accepted
Published
Volume
16 - 2026

Copyright

Copyright © 2026 Mulla, Balbaid, Aldosary, Ghandourh, AlDahlawi, Bayoumy, Alhumaid, Alrowais, Almerdhemah, Taha, AlMohamad and Jastaniah.

License: https://creativecommons.org/licenses/by/4.0/

*Correspondence: Zaheeda Mulla

Article metrics

Views
0
Downloads
0
Citations
0
Shares
0