Diagnostic accuracy of rapid and point-of-care tests for Schistosoma haematobium in African populations: a systematic review and meta-analysis

  • Erica Draxl1*corresponding author
  • Eleanor Ochodo2,3
  • Pavitra Pillay4
  • Ntombizodwa Ndlovu1
  1. 1School of Public Health, Faculty of Health Sciences, University of Witwatersrand, Johannesburg, South Africa
  2. 2Centre for Global Health Research, Kenya Medical Research Institute, Kisumu, Kenya
  3. 3Centre for Evidence-Based Health Care, Department of Global Health, Faculty of Medicine and Health Sciences, Stellenbosch University, Stellenbosch, South Africa
  4. 4Department of Biomedical and Clinical Technology, Faculty of Health Sciences, Durban University of Technology, Durban, South Africa
Article metrics0 viewsView details

Abstract

Accurate and accessible point-of-care testing for urogenital schistosomiasis (UGS) is vital for control and prevention in sub-Saharan Africa. Point-of-care (POC) rapid diagnostic tests with high sensitivity for both high and low parasite burdens have been developed. These tests need to be assessed for their applicability in poorly resourced UGS-endemic areas so that control programs can be initiated, and improved screening and routine diagnosis can be implemented. This diagnostic test accuracy (DTA) review evaluates the sensitivity and specificity of eligible field-based index tests compared to laboratory-based reference tests for the diagnosis of UGS among individuals from Africa. Four databases were systematically searched for DTA studies on S. haematobium -exposed patients in endemic settings from 2000 to 2026. Methodological quality was assessed using the QUADAS-2 tool. Using the variability of test thresholds, a bivariate or HSROC model was used to calculate the diagnostic performance for each eligible test. Heterogeneity and subgroup analyses were conducted for the following variables: reference-standard comparators, age groups, and prevalence settings. Twenty-nine studies were evaluated, in which nine different POC and rapid tests were eligible, and four were included in a meta-analysis. Two field-based molecular tests, loop-mediated isothermal amplification (LAMP) and recombinase polymerase amplification (RPA), demonstrated high accuracy (sensitivity, 94.7%; specificity, 66.4%; and sensitivity, 94.1%; specificity, 96.6%, respectively). The immunochromatographic-rapid diagnostic test (ICT-RDT) and S. mansoni cercarial transformation fluid RDT (SmCTF-RDT) serological tests are easy-to-use POC cassettes and have high sensitivity (98.6% and 98.4%, respectively) but require a portable centrifuge to obtain serum samples. Other serological tests: dipstick dye immunoassay (DDIA) and indirect haemagglutination assay (IHA) had lower sensitivities (64.8% and 59.5%, respectively). The POC circulating cathodic antigen test (POC-CCA) demonstrated a sensitivity of 49.8% and specificity of 77.0% and has low diagnostic accuracy for S. haematobium. Portable and mobile-phone microscopes (sensitivity, 81.5%, specificity, 91.4%) were less sensitive than conventional microscopy but can have benefits in field-based low-resource laboratories. Further validation with improved reference standards is required to assess the diagnostic accuracy of many of the tests identified. RPA and LAMP emerged as promising field-applicable rapid tests for schistosomiasis control.

Keywords

Introduction

Schistosomiasis, or bilharzia, is a water-borne neglected tropical disease caused by the parasitic blood flukes of the genus Schistosoma. Schistosomiasis affects more than 150 million people worldwide, 84.3% of whom live in sub-Saharan Africa (SSA) (Li et al., 2025). Two types of Schistosoma species exist in SSA: Schistosoma haematobium, which causes urogenital schistosomiasis, and Schistosoma mansoni, which causes intestinal schistosomiasis (Gray et al., 2011). Both are acquired by exposure to infected water harboring specific snail species that serve as intermediate hosts for Schistosoma species (Adenowo et al., 2015). Intestinal schistosomiasis results in lesions in the intestinal mucosa and liver, which can result in bleeding, malnutrition, advanced disease manifestations, and even death due to fibrosis in the portal veins of the liver. Urogenital schistosomiasis results in pathology causing damage to the bladder, kidneys, and genital organs, leading to symptoms such as painful urination and hematuria, and can advance to bladder cancer or obstructive uropathy (Gray et al., 2011).

Chronic or long-standing schistosomiasis results in lifelong debility due to malnutrition, internal hemorrhage, and end-stage organ failure, contributing to 1.746 million disability adjusted life years (DALY’s) lost annually (Shen and Luo, 2025). S. haematobium infection in the genital tract is called female or male genital schistosomiasis (FGS or MGS, respectively) and results in an increased risk of acquiring sexually transmitted diseases, including HIV (Sturt et al., 2020). Numerous studies in SSA have ascertained that FGS is associated with a 2.9 to 4-times higher risk of HIV acquisition (Sturt et al., 2020).

Current schistosomiasis control and elimination strategies rely on periodic mass drug administration (MDA) with a global target of regular treatment of at least 75% of school-aged children at risk (World Health Organization, 2022). Effective and efficient MDA requires adequate surveillance (Stothard et al., 2009), particularly for schistosomiasis-endemic regions. However, only half of the target populations in endemic SSA countries currently receive MDA (Hong, 2023), and many countries are unable to initiate control programs due to challenges in implementation (Randjelovic et al., 2015). Increasingly, the need for improved diagnostic tools has been recognized by the World Health Organization (WHO) and schistosomiasis research communities as an imperative first step to improve control strategies, and for improved geographic mapping and prevalence in endemic areas (Hong, 2023). Fast, reliable, and near-patient diagnostic tests are an integral part of sustained control programs and decision-making for schistosomiasis elimination programs.

There is a notable gap in knowledge regarding diagnostic tests designed to detect S. haematobium in low-resource settings, where access to laboratory equipment is limited. Although microscopy of filtered urine for egg detection is considered the reference or “gold standard” diagnostic method, it has significant limitations (Hong, 2023). The sensitivity of microscopy is reliable primarily where high infection intensity occurs endemically or in ‘hotspots’ where infection rates exceed 50% (World Health Organization, 2022). However, in low-prevalence regions and among individuals with chronic infections, its sensitivity is markedly reduced, hindering early diagnosis and timely treatment (Hong, 2023). For instance, in women with chronic schistosomiasis diagnosed with female genital schistosomiasis (FGS), only 23% were found to have detectable eggs in urine samples using microscopy (Poggensee et al., 1998). This discrepancy highlights the urgent need for more sensitive diagnostic tools to replace conventional parasitological methods, particularly for accurate prevalence assessments and improved case detection.

The ideal test for resource-limited settings would be a direct point-of-care (POC) test conducted at or near the clinical site that is rapid, easy to use, and highly sensitive and specific for all Schistosoma species. An appropriate rapid turnaround time between performing the test and obtaining the result is 10 minutes to 2 hours (World Health Organization, 2014). Such a test would support timely decision-making and improve patient management (Drain et al., 2014).

Direct diagnostic methods include antigen tests and molecular (DNA-based tests), all of which identify the schistosome parasite itself and are therefore considered more accurate for prevalence estimation, surveillance, and case detection (Chala, 2023). In contrast, indirect tests do not detect the parasite directly. These include urine reagent strips and antibody-based serological tests, which serve as proxy measures of infection and are often used for screening. Although serological tests are indirect, they remain clinically valuable because they detect antibodies, which are the host’s immune response to Schistosoma infection. Anti-Schistosoma antibodies may persist for life in untreated infections. For this reason, serology can support the diagnosis of asymptomatic or untreated chronic schistosomiasis, particularly when eggs or other antigens are difficult to detect by other methods. After treatment with praziquantel, anti-Schistosoma antibodies may take approximately 72–100 weeks to become undetectable (González-Sanz et al., 2025).

Serological tests are therefore useful as rapid, low-cost screening tools in individuals with untreated chronic infections, as well as selected newly infected populations, including recent travellers or migrants returning from endemic areas (González-Sanz et al., 2025) and young children with new infections (Wami et al., 2014).

For real field applications demanded by resource-limited settings, the ASSURED criteria (Affordable, Sensitive, Specific, User-friendly, Robust & Rapid, Equipment-free, Deliverable) for point-of-care diagnostic tests were developed by the WHO in 2003 (Mabey et al., 2004), and updated to REASSURED in 2019, where Real-time connectivity and Ease of specimen collection were included in the diagnostic framework (Land et al., 2019). Currently, the commercially available easy-to-use lateral flow POC-CCA test is considered accurate for S. mansoni and meets ASSURED criteria (Danso-Appiah et al., 2016), however its accuracy is low for S. haematobium, making it inaccurate for areas of co-endemicity (Ochodo et al., 2015; Rubaba et al., 2018).

Despite the need, the optimal approach for rapid diagnosis of urogenital schistosomiasis has not been fully established. A recent diagnostic test accuracy (DTA) review in the literature assessing tests for both S. haematobium and S. mansoni found no viable cost-effective practical alternatives to conventional microscopy for S. haematobium-prevalent areas (Vaillant et al., 2024). There is a need to synthesize the available evidence on POC and rapid tests for the rapid diagnosis of S. haematobium, particularly in low-resource settings with varying levels of disease prevalence and chronicity. Recent advancements in the development of low-cost, field-based versions of existing expensive laboratory-based tests also warrant attention, as they hold promise for enhancing surveillance and control efforts in endemic and non-endemic areas (Ally et al., 2024).

In this systematic review, we assessed and compared the diagnostic accuracy of eligible rapid and point-of-care tests with a timing of approximately 2 hours for the detection of S. haematobium infection, using appropriate laboratory tests as the reference standard in exposed African populations. We also compared the diagnostic accuracy of eligible tests in different areas of prevalence (endemic high/moderate/low), age-groups (children and adults), and for different reference tests (conventional microscopy, PCR, composite tests) using subgroup analysis. This review aims to identify appropriate POC or rapid tests that can serve as a suitable replacement for conventional microscopy in the formulation of integrated control strategies in low-resource settings.

Methods

This systematic review of diagnostic test accuracy (DTA) followed the Preferred Reporting Items for Systematic Reviews and Meta-Analyses-Diagnostic Test Accuracy (PRISMA-DTA) guidelines (Mcinnes et al., 2018) and the Cochrane DTA handbook (Dinnes et al., 2022).

Search strategy and study selection

Four major databases, PubMed, Global Health, ProQuest, and the Web of Science, were systematically searched for eligible DTA studies published between January 2000 and June 2026. The medical subject headings (MeSH) search terms used were (Schistosoma haematobium) AND (diagnostic test) AND (point of care test) AND (diagnostic accuracy) OR (Urogenital schistosomiasis) AND (diagnostic test) OR (Schistosoma haematobium) AND (rapid diagnostic test) OR (Diagnostic accuracy) AND (Schistosoma haematobium). A second search strategy and article mining were conducted for individual index tests after the initial search to ensure that the totality of diagnostic performance studies was included. An example of search terms used was: (Schistosoma haematobium) AND (“LAMP”), depending on each different index test. Article mining in the references list of relevant reviews was also performed to ensure all studies were included in the search (Feleke et al., 2023) (Vaillant et al., 2024), (Zhou et al., 2025).

Following the removal of duplicates, two reviewers independently screened the remaining studies. Exclusions were done by title, then by abstract, and then by full-text screening, based on pre-determined inclusion/exclusion criteria. Any studies that had discrepancies between the reviewers were screened and resolved by consensus, or if necessary, independently resolved by a third reviewer. Suitable studies were uploaded into the Rayyan reference manager software (www.rayyan.ai) for the ease of independent review.

Eligibility criteria of included studies

Studies were included if they analysed the diagnostic accuracy of POC or rapid field-applicable tests for the diagnosis of S. haematobium infections in individuals of all ages from endemic African countries. Non-species-specific indirect immunological tests were also included if the study used a reference test that diagnosed S. haematobium to confirm the diagnosis of UGS. Only studies in the English language were included in this systematic review and meta-analysis. Information about sensitivity and specificity or true-negatives (TNs), true-positives (TPs), false-positives (FPs), and false-negatives (FNs) was calculated from the data presented to produce contingency tables for diagnostic accuracy and receiver operating characteristic (ROC) analyses. Studies where this data could not be retrieved from the text were excluded. Trace positives were not considered in our review and were excluded from the accuracy calculations where possible. Primary observational studies were included if the results of an index test were compared to that of an appropriate laboratory-based reference standard. Case-control studies were included if cases and controls were sampled from the same patient population but were excluded if the control was from a healthy population from a non-endemic area. Proof-of-concept studies using artificially spiked urine were excluded, as well as commentaries, case studies, or editorials.

To provide an overall estimate of diagnostic accuracy, various laboratory-based reference tests were eligible, including standard microscopy, composite tests, PCR, or statistical models using Latent Class Analysis (LCA). Data about the reference test included sample preparation and the diagnostic threshold used to report a positive result. Bayesian modelling or any reference method where no test on biological samples was performed was excluded. In selecting eligible reference tests, parasitological methods were included, where urine sample concentration techniques (sedimentation, filtration, or centrifugation) and the use of single or double/triple microscopy were reported. Composite reference tests and LCA modelling using combined results from more than one test were eligible, provided that the type of tests used as comparators was reported. Studies that used the index test results as part of the composite reference test were excluded. For molecular-based reference tests, different types of PCR were accepted, including real-time PCR, the Dra1-based qPCR, and conventional PCR, provided that the DNA extraction and amplification techniques were reported. ELISA, Western Blot, or any laboratory test using biological samples from patients was included.

In selecting eligible index tests, POC tests were defined as those that can be conducted at or near the clinical site to facilitate timely and cost-effective decision-making (Drain et al., 2014). Rapid tests were defined as field versions of laboratory tests that use reagents that are not heat sensitive, utilising portable table-top equipment (such as mini centrifuging devices), and still ensure results are available within 2½ hours (World Health Organization, 2014). Relevant tests that meet the eligibility criteria of direct tests are antigen tests, molecular tests, and parasitological tests. The diagnostic methods include detecting Schistosoma antigens, schistosome DNA from the parasite itself, or viewing the eggs of S. haematobium in filtered urine. Rapid serological tests are also considered eligible for this review, as indirect tests, as their diagnostic value in new infections (Wami et al., 2014), and chronic infections is noted (González-Sanz et al., 2025). Urine reagent strips for haematuria and proteinuria have been excluded from this review as these tests serve as proxy tests and are unable to directly diagnose S. haematobium.

Data extraction

Key characteristics of each study were extracted and summarised in a data extraction table in Microsoft Excel (Microsoft Corporation, 2022). The data extracted included study authors, publication year, journal name, country, prevalence of schistosomiasis in the population (low (<10%)/moderate (10%-50%)/high (>50%)), participant characteristics (including age, sex and method of participant selection), reference test standard used, index test (including name of manufacturer or equipment used, DNA extraction and amplification methods for molecular tests), index test sample, and threshold for positive results. The numbers of TPs, FNs, FPs, and FNs were entered into a 2 x 2 contingency table to calculate sensitivity and specificity measures manually, and these were compared to the published results for each study.

Quality assessment

The methodological quality of the eligible studies was reviewed using a critical appraisal tool (Quality Assessment of Diagnostic Accuracy Studies 2, QUADAS-2). The tool was used to assess the risk of bias for diagnostic accuracy studies (Whiting et al., 2011). Four key domains of risk of bias and applicability to the research question assessed were patient selection, reference standard, index test, and flow and timing.

Statistical analysis

From the data extraction table, studies were grouped according to index tests. Raw data from the extraction process were imported into the RevMan 5.3 software (training.cochrane.org) as recommended by the Cochrane Screening and Diagnostic Test Methods Group (SDTM) (Macaskill et al., 2010), to generate forest plots and ROC curves. TPs s, FNs, FPs, and FNs were entered into RevMan to calculate sensitivity and specificity measures with 95% confidence intervals for each study. A forest plot was generated for each index test group to visually report the sensitivity and specificity measures and confidence intervals. A ROC plot was generated from the data, producing a summary curve to visualise how the sensitivity and specificity measures of the tests performed against each other. The forest plots and ROC curves for each index test group were analysed graphically for heterogeneity, which was classified as low, medium, or high.

A meta-analysis was performed for all index tests appearing in three or more publications and which had no substantial heterogeneity. Tests with fewer than three studies were analysed narratively. Pooled estimates of sensitivity and specificity for each index test were calculated in STATA version 17 (STATA Corporation, 2015 statistical software: Release 14. College Station, Texas: StataCorp LP) or SAS version 9.3. (Copyright 2024 SAS Institute INC). The statistical model selected for the overall analysis for each test depended on how studies chose thresholds to determine positivity.

For tests that have a single common threshold, the bivariate random-effects model, as proposed by Reitsma was used for overall meta-analysis (Reitsma et al., 2005). The user-written command ‘metandi’ was utilized in STATA to fit the model and achieve pooled diagnostic accuracy results (Harbord and Whiting, 2009). A hierarchical summary (HSROC) model was utilised for index tests with variability among studies due to multiple positivity thresholds (Rutter and Gatsonis, 2001). HSROC provides an average diagnostic accuracy result to account for the within- and between-study variability among studies. The meta-analysis provided the final diagnostic accuracy result for each index test group.

Investigations for heterogeneity

An exploratory analysis was performed for index test groups that visually displayed high heterogeneity in the forest plots and ROC analyses. An I2 statistic was not utilized due to the high variability in index and reference tests (Naaktgeboren et al., 2016). To determine if a factor/co-variate was associated with heterogeneity in test accuracy, forest plots and ROC curves were created in RevMan 5.3 for the following variables: prevalence setting, age-group, and reference test. These variables were chosen as possible reasons for heterogeneity according to the information provided in each study. ROC curves were generated for each of these three variables within each index test group and were examined visually to assess variations in accuracy for each variable. The heterogeneity was visually assessed as high, moderate, or low.

Index tests displaying high heterogeneity were further analysed using sub-group analysis techniques in STATA to determine actual sensitivity and specificity measures for each sub-group. For tests analysed with the bivariate random-effects model, we calculated sensitivity and specificity measures for each subgroup in STATA and then compared these results with the total pooled sensitivity and specificity measures. For tests analysed with the HSROC model, we investigated whether these covariates graphically affect the shape of the curve of the ROC plot generated separately for each covariate. Comparisons were made between the overall accuracy and subgroup accuracy for each index test.

Results

Search results and eligible studies

The search of databases identified 593 peer-reviewed studies on the diagnostic performance of POC/rapid tests for S. haematobium. A second index test-specific search produced 93 additional studies and 4 studies from citation screening in relevant reviews. After removing 110 duplicates, a total of 580 studies were screened independently by two reviewers who identified and excluded 378 non-eligible studies through title and abstract screening. A total of 202 studies were retained and assessed for eligibility. Twenty-nine studies were eligible according to our inclusion criteria, and data were extracted from these. Four studies analysed the diagnostic performance of more than one test; therefore, 36 evaluations from 29 studies were included. The 29 studies represent 9943 participants. The reasons for exclusion are reported in the PRISMA flowchart of the study selection process in Figure 1.

PRISMA flow diagram for the identification of studies included in this systematic review.
Figure 1. PRISMA flow diagram for the identification of studies included in this systematic review.

Characteristics of included studies

The characteristics of the 29 included studies are presented in Table 1. All the studies included in the review were published from January 2000 to June 2026. Most of the studies, 26 (89.6%), were in field settings (village, school, or rural district) where study participants were from the general population in endemic areas. Two studies used banked samples from previous studies, and one study included patients in an outpatient clinic with suspected UGS.

Table 1. Characteristics of studies included in the systematic review and meta-analysis.

Author, yearCountryParticipantsStudy designSettingPrevalence ofS. haematobiumBiological sampleTargetIndex test (Thresholds)Reference test
Serological tests
(Yameny, 2019)EgyptAdults - 50 with UGS positive and 50 with UGS negative samples, age not statedCase controlEndemicModerate (15%)Sh prevalentSerumAdult Worm AntigenIHA (cutoff titre >1.160)Single microscopy of centrifuged, filtered, or sedimented urine
(Zhang et al., 2020)Zambia148 school-aged children aged 7 to 15Cross sectionalEndemicHigh (61%)Co-endemic Sh/SmSerumAdult Worm AntigenSoluble Egg AntigenIHA (cutoff titre >1:1.20)DDIA (colour)Double urine microscopy on filtered urine
(Houmsou et al., 2021)Nigeria374 participants from a rural area in Nigeria, all agesCross sectionalEndemicModerate 33%Sh prevalent areaSerumAdult Worm AntigenICT-RDT (+test band)Single microscopy of filtered urine
(Mudenda et al., 2022)Zambia430 adults residing in Siyavonga district, 18–50 yearsCross sectionalEndemicLow 4%Co-endemic Sh/SmSerumAdult Worm AntigenICT-RDT (+test band)Single microscopy of filtered urine
(Nausch et al., 2014)Zimbabwe91 school children with egg counts for S. mansoni and/or S. haematobium, age not statedCross sectionalEndemicHigh, 58%Sh prevalent areaSerumCercarial transformation fluid (SmCTF)SmCFT-RDT (+test band)Single microscopy of filtered urine
(Coulibaly et al., 2013)Cote Ivoire118 preschool children (3–6 years)Cross sectionalEndemicLow, 5.1%Co-endemic Sh/SmFinger-prick bloodCercarial transformation fluid (SmCTF)SmCFT-RDT (+test band)Double microscopy of filtered urine
Antigen Tests
(Stothard et al., 2006)Zanzibar Tanzania50 school children - 25 positive, 25 negative for S. haematobium, age not statedCase controlEndemicHigh, 73%Sh prevalent areaUrineCathodic antigenPOC-CCA (+test band)Single microscopy of filtered urine
(Obeng et al., 2008)Ghana152 school children, 5–16 yearsCase controlEndemicLow, 0-20%Co-endemic Sh/SmUrineCathodic antigenPOC-CCA (+test band)Single microscopy of filtered urine
(Ayele et al., 2008)Ethiopia206 school children, 4–21 yearsCross sectionalEndemicHigh 47%Sh prevalent areaUrineCathodic antigenPOC-CCA (+test band)Single microscopy of filtered urine
(Ashton et al., 2011)Sudan373 children, 5–16 yearsCross sectionalEndemicModerate, 26%Co-endemic Sh/SmUrineCathodic antigenPOC-CCA (+test band)Single microscopy of filtered urine
(El-Ghareeb et al., 2016)Egypt96 school children, 5–12 yearsCross sectionalEndemic Post MDALow, 5.3%Sh prevalent areaUrineCathodic antigenPOC-CCA (+test band)Single microscopy of filtered urine
(Sanneh et al., 2017)The Gambia1954 school children, 7–14 yearsCross sectionalEndemicLow, 10%Sh prevalent areaUrineCathodic antigenPOC-CCA (+test band)Single microscopy of filtered urine
(Rubaba et al., 2018)South Africa420 school children, 10–15 yearsCross sectionalEndemicModerate, 40%Sh prevalent areaUrineCathodic antigenPOC-CCA (+test band)Single microscopy of filtered urine
(Robinson et al., 2021)Madagascar500 adults and children, age not statedCross sectionalEndemicHigh, 47%Co-endemic Sh/SmUrineCathodic antigenPOC-CCA (+test band)Single microscopy of filtered urine
(Sheele et al., 2013)Kenya160 school children with recent blood in urine, 8–17 yearsCross sectionalEndemicHigh 49%Sh prevalent areaUrineEggsRDT-Sh (colour change)Single microscopy of filtered urine
Nucleic acid amplification tests (NAAT’s)
(Gandasegui et al., 2018) (b)Angola172 school children, aged 8–15 yearsCross sectionalEndemicHigh, 50.6%Sh prevalent areaUrineDNA (Ribosomal intergenic spacer)Sh-LAMPConventional method(fluorescence)Single microscopy of filtered urine
(Gandasegui et al., 2018) (a)Angola172 school children, aged 8–15 yearsCross sectionalEndemicHigh, 50.6%Sh prevalent areaUrineDNA (Ribosomal intergenic spacer)Sh-LAMPRapid-Heat LAMPellet method(fluorescence)Single microscopy of filtered urine
(Bayoumi et al., 2016)Egypt69 adults with suspected schistosomiasis at an outpatient clinicCross sectionalEndemicModerate, 44.9%Co-endemic Sh/SmUrineDNA (Ribosomal intergenic spacer)Sh-LAMPConventional method (fluorescence)Single microscopy of filtered and centrifuged urine
(Lodh et al., 2017)Ghana86 participants, aged 5–26 yearsCross sectionalEndemicHigh, 81%Co-endemic Sh/SmUrineDNA (Ribosomal intergenic spacer)Sh-LAMPConventional method(fluorescence)PCR Urine
(van Bergen et al., 2024)MadagascarArchived DNA samples extracted from 125 vaginal mucosal swabs from women aged 15–35 years.Cross SectionalEndemicHigh, 67%Sh prevalent areaCervico-vaginal fluidDNA (Ribosomal intergenic spacer)Sh-LAMPConventional method(fluorescence)qPCR of cervical lavage samples
(Archer et al., 2022)Zambia527 non-pregnant, sexually active women, 18–35 years.Cross sectionalEndemicLow, 2.5%Sh prevalent areaCervico-vaginal fluidDNA (Sh-Dra1 repeat sequence)RPA-Sh (200mV)PCR of cervical lavage samples
(Frimpong et al., 2021)Ghana1290 participants resident in Kumasi, from whom 69 positive and 66 negative urine samples were chosen, all agesCase ControlEndemicLow, 5.3%Co-endemic Sh/SmUrineDNA (Sh-Dra1 repeat sequence)RPA-Sh (900mV)Composite reference (RT-PCR of urine and single microscopy of centrifuged urine)
(Archer et al., 2020)Zanzibar Tanzania168 school children - 100 UGS positive and 68 random controls, 9–12 yearsCase controlEndemicLow, <5%Sh prevalent elimination settingUrineDNA (Sh-Dra1 repeat sequence)RPA-Sh (200mV)Single microscopy of filtered urine
Parasitological tests
(Armstrong et al., 2022)Ghana and Côte d’Ivoire205 School-aged children, of which 63 were from Ghana and 142 from Côte d’IvoireCross sectionalEndemicHigh 55%Sh prevalent areaUrineEggsPortable microscope: Schistoscope AIConventional single microscopy of filtered urine
(Meulah et al., 2024) (a)Gabon339 school-aged children (aged 7-13) and adultsCross sectionalEndemicModerate 30%Sh prevalent areaUrineEggsPortable microscope: Schistoscope AIConventional single microscopy of filtered urine
(Meulah et al., 2024) (b)Gabon339 school-aged children (aged 7-13) and adultsCross SectionalEndemicModerate 30%Sh prevalent areaUrineEggsPortable microscope: Schistoscope AIComposite reference(Single microscopy, rt-PCR urine, UCP-LFCAA urine)
(Meulah et al., 2024) (c)GabonBiobanked urine samples from 798 school-aged children, adults and pregnant womenCross SectionalEndemicModerate 30%Sh prevalent areaUrineEggsPortable microscope: Schistoscope AIConventional single microscopy of filtered urine
(Meulah et al., 2024) (d)GabonBiobanked urine samples from 349 school-aged children, adults and pregnant womenCross sectionalEndemicModerate 30%Sh prevalent areaUrineEggsPortable microscope: Schistoscope AIConventional single microscopy of filtered urine
(Meulah et al., 2024) (e)GabonBiobanked urine samples from 349 school-aged children, adults and pregnant womenCross SectionalEndemicModerate 30%Sh prevalent areaUrineEggsPortable microscope: Schistoscope AIComposite reference(Single microscopy, rt-PCR urine, UCP-LFCAA urine)
(Meulah et al., 2022)Nigeria487 pre-school, school-aged children, and adultsCross sectionalEndemicHigh 49%Sh prevalent areaUrineEggsPortable microscope: Schistoscope automated 5.0Conventionalsingle microscopy of homogenised filtered urine
(Bogoch et al., 2017)Ghana223 school children who provided 60 urine samples, 7–16 yearsCross sectionalEndemicHigh, 73%Sh prevalent areaUrineEggsMobile-phone handheldmicroscopeConventional single microscopy of centrifuged urine
(Ephraim et al., 2015)GhanaSchool children who provided 49 urine samples, 7–13 yearsCross sectionalEndemicHigh, 69%Sh prevalent areaUrineEggsMobile-phone handheldmicroscope(Cellscope)Conventional single microscopy of centrifuged urine
(Coulibaly et al., 2016) (a)Cote d’Ivoire226 children and young adults aged 6–19 yearsCross sectionalEndemicModerate, 40%Co-endemic Sh/SmUrineEggsMobile-phone handheld microscope (Newton NM1)Conventional single microscopy of filtered urine
(Coulibaly et al., 2016) (b)Cote d’Ivoire226 children and young adults aged 6–19 years.Cross sectionalEndemicModerate, 40%Co-endemic Sh/SmUrineEggsMobile-phone handheld microscope (Cellscope)Conventional single microscopy of filtered urine
(Coulibaly et al., 2023)Cote d’Ivoire170 children aged 5–14 yearsCross sectionalEndemicModerate 20%Co-endemic Sh/SmUrineEggsPortable microscope: Schistoscope AIConventional single microscopy of filtered urine

Sh: S.haematobium, Sm: S.mansoni.

The median prevalence of S. haematobium infection was 30.7% (range 2.5% - 73.0%). In Table 1, prevalence is defined as high (infection diagnosed in >51% of school-aged children (SAC)), moderate (>10% and <50% in SAC), and low (<10% in SAC). In 6 studies (20.6%), participants in co-endemic populations were tested for both S. haematobium and S. mansoni using non–species-specific antigen or antibody assays, although the results have been presented for S. haematobium only. Fourteen studies focused on diagnostic tests that specifically identify evidence of S. haematobium for the diagnosis of UGS only. Most of the studies were cross-sectional or cohort studies (86.2%), either using urine/serum samples from previous studies or collecting urine samples from participants in field studies. Four were case-control studies comparing serum samples from individuals with UGS and without UGS from the same population.

Nine rapid diagnostic testing techniques were identified as eligible index tests through the systematic search for diagnostic accuracy studies, as presented in Table 2. Four of these index tests were entered into the meta-analysis, as five index tests had two or less diagnostic accuracy studies published and therefore were analysed narratively. Only the circulating cathodic antigen test (POC-CCA) is considered a POC test because it can be conducted at or near the clinical site (Drain et al., 2014). The other eight tests are considered rapid as they use field-based kits and minimal portable laboratory equipment, with results available within 2½ hours.

Table 2. Diagnostic tests identified as eligible rapid and POC index tests through a systematic search of diagnostic accuracy studies and included in this review.

TestTest name and descriptionTargetRapid/POCNo. of reportsReason for consideration
LAMPLoop Mediated Isothermal AmplificationLAMP is a nucleic acid amplification test (NAAT) that amplifies schistosome DNA under isothermal conditions (60–65 °C)Sample required: UrineSpecies specific for S. haematobium.DNA(Ribosomal intergenic spacer)Rapid5Time to result: 10 mins for DNA extraction, 1–2 hours for DNA amplification (Gandasegui et al., 2018).Field applicability: Minimal equipment needed, centrifuge, heat block, commercial LAMP reagents
Sh-RPARecombinase polymerase amplificationThe Sh-RPA assay is a NAAT that amplifies schistosome DNA at constant isothermal temperatures (37-42 °C).Sample required: UrineSpecies specific for S. haematobiumDNA(Sh-Dra1 repeat sequence)Rapid3Time to result: Approximately 1 hour: 10 mins for DNA extraction and amplification, 15 mins for RPA reaction at 42°C, 10 mins for Fluorometer reading (Archer et al., 2022)Field applicability: Minimal equipment needed; the RPA reaction is initiated in test tubes with hand shaking or centrifuging. A portable testing device measures fluorescence.
IHAIndirect haemagglutination assaySerological test detecting antibodies in serum to worm antigen using haemagglutination techniques.Sample required: SerumSpecies specific: No. Cannot distinguish between S. haematobium and S. mansoni.Adult Worm AntigenRapid2Time to result: Approx 2 ½ hours:10 minutes for obtaining serum, 10 minutes for the agglutination step, 2 hours for incubation and result (Yameny, 2019).Field applicability: Simple test requiring only basic laboratory equipment: table-top centrifuge, reagents from the IHAT Fumouze kit.
ICT-RDTImmunochromatographic- rapid diagnostic testSerological test detecting antibodies to Schistosoma spp antigen in serum, using a lateral flow cassette.Sample required: SerumSpecies specific: No. Cannot distinguish between S. haematobium and S. mansoni.Adult Worm AntigenRapid2Time to result: 20–30 minutes.10 minutes for obtaining serum, 10 minutes for the test (Mudenda et al., 2022).Field applicability: simple to use, only a tabletop centrifuge is required to attain serum
SmCTF-RDTS.mansoni-cercarial transformation fluid- RDTSerological test detecting antibodies to S.mansoni cercarial transformation fluid (SmCTF) in serum.Sample required: Serum or finger-prick bloodSpecies specific: No. Cannot distinguish between S. haematobium and S. mansoni.Cercarial transformation fluidRapid2Time to result: 15 to 20 minutes.10 minutes to obtain serum, 10 minutes for the test (Nausch et al., 2014).Field applicability: simple to use, only a tabletop centrifuge is required to attain serum
DDIADipstick Dye Immuno AssaySerological test detecting antibodies to soluble egg antigen (SEA) in serum.Sample required: SerumSpecies specific: No. Cannot distinguish between S. haematobium and S. mansoni.Soluble Egg AntigenRapid1Time to result: 20 minutes.Field applicability: A table-top centrifuge is required to attain serum,PVC well, micropipette, dipstick (Zhang et al., 2020).
RDT-ShRapid Diagnostic Test- S. haematobiumAn antigen test designed to detect eggs in urine by antibody-antigen reaction on the egg surface.Sample required: UrineSpecies specific for S. haematobiumEggsRapid1Time to result: 20 minutesField applicability: Yes, simple and cheap, no laboratory needed, only a syringe and filter coupling device, reagent, and cup (Sheele et al., 2013)
POC-CCAPoint of Care- Circulating Cathodic antigenUrine- based lateral flow antigen test cassette that detects the circulating cathodic antigen (CCA) from the adult Schistosoma worm in urine,Sample required: UrineSpecies specific: No. Cannot distinguish between S. haematobium and S. mansoni.Cathodic AntigenPOC8Time to result: 20 minutes.Field applicability: Yes, simple to use; 1 drop of urine and a subsequent drop of buffer (old test version only). No laboratory equipment needed
Portable/mobile phone microscopyThree portable microscopes using a mobile phone or digital/AI technology are included:-The Cellscope: attached to a mobile phone- The Newton 1: attached to a mobile phone-The Schistoscope, which is a digital automated microscope, is attached to a laptop.Sample required: UrineSpecies specific for S. haematobiumEggsRapid12Time to result: 45 mins to 1 hourField applicability: The devices can be used without laboratory equipment as stand-alone devices
TOTAL36

The indirect haemagglutination assay (IHA) and the immuno-chromatographic rapid diagnostic tests (ICT-RDT), the S. mansoni-cercarial transformation fluid-RDT (SmCTF-RDT), and the dipstick dye immunoassay (DDIA) are non-species-specific serological screening tests that detect antibodies in serum (Yameny, 2019) (Houmsou et al., 2021) (Nausch et al., 2014). These assays require blood specimens and the availability of portable centrifuges to obtain serum in the field. The POC-CCA detects circulating cathodic antigen (CCA) in urine but is also not species-specific, with rapid results in under 30 minutes using a simple urine-based lateral flow cassette. Non-species-specific tests are excellent screening tests for both urinary and intestinal schistosomiasis. However, in areas where both species exist, these tests need to be used with caution if the sensitivity of the test for either species differs, to prevent false negatives. The POC-CCA is known to diagnose S. mansoni, but has poor accuracy for S. haematobium (Ochodo et al., 2015; Rubaba et al., 2018), resulting in underdiagnosis in co-endemic areas. Eleven included studies (37.9%) were performed in areas where both S. haematobium and S. mansoni were prevalent, or in populations from co-endemic areas, further highlighting the need for species-specific tests, or tests with high accuracy for both species of Schistosoma.

One study, the RDT- S. haematobium (RDT-Sh), describes a novel, cheap, and simple method to detect antibodies in urine samples using a syringe, filter, and anti-IgG for detecting schistosome eggs specific for S. haematobium (Sheele et al., 2013).

With recent advances in molecular DNA-based field-applicable testing, the loop-mediated isothermal amplification (LAMP) and the recombinase polymerase amplification (Sh-RPA) tests are both being developed to be rapid diagnostic assays to detect schistosome DNA fragments in urine samples. The LAMP and RPA tests are Schistosoma species-specific tests that can distinguish between S. haematobium and S. mansoni. In these tests, the DNA extraction and amplification reagents used do not require cold chain logistics and minimal equipment is required (Gandasegui et al., 2015) (Archer et al., 2020). DNA extraction requires centrifuging of urine to obtain a pellet of urinary sediment, and the amplification reaction requires incubation at 65°C using a heating block. Both tests are rapid tests using minimal laboratory equipment, with results obtainable within 2½ hours in field settings.

Portable/mobile-phone microscopy using portable devices connected to a mobile phone are also species-specific tests that can identify S. haematobium eggs in filtered or centrifuged urine. The Schistoscope was categorized with the portable/mobile-phone microscopes, as this is an automated microscope developed specifically for low-resource settings (Meulah et al., 2022).

In Table 1, test thresholds to determine positivity are provided in brackets next to each index test so that variable test thresholds can be identified for each index test group. Lateral flow cassette tests which have a colour change for positive results are indicated as + test band. Tests are categorized as serological, antigen, NAAT (molecular), and parasitological.

It can be noted that studies conducted before 2020 used single microscopy as a reference test, and those after 2020 used improved reference tests such as PCR, latent class analysis, or double microscopy. In high-prevalence endemic areas, microscopy can be considered an appropriate reference test (World Health Organization, 2022). However, 6 studies (20.6%) used microscopy of urine as a reference test in low/moderate prevalence settings, which could cause bias in the overall results.

Appraisal of risk of bias and applicability

The QUADAS-2 risk assessment results are presented in Figure 2 and further elaborated in Supplementary Additional File 2. The overall methodological quality of the studies was rated as moderate. Methods of participant selection were adequately described; however, eight studies (25%) chose known positive and negative samples instead of randomly selected samples and were therefore classed as high risk of bias for patient selection. Selection bias can be described in one of the studies, where participants were chosen from patients who attended health clinics and were not representative of the general population. In over 75% of the studies, random selection of participants was conducted.

Risk of bias and applicability concerns for the studies reviewed (Note: the authors’ judgements about each domain are presented as percentages across included studies.).
Figure 2. Risk of bias and applicability concerns for the studies reviewed (Note: the authors’ judgements about each domain are presented as percentages across included studies.).

Over 85% of the studies appropriately reported the index test, described how the tests were conducted, ensured that results from each test were blinded to prevent bias, and explained how the results were interpreted. In the index test domain, risk of bias relates to tests that are not blinded. The accuracy of the reference tests differed by reference test type and prevalence of infection. In the reference test domain, more than 40% of studies in low and moderate prevalence settings used the inaccurate single microscopy test as the reference, resulting in a high risk of bias. In the flow and timing domain, approximately 10% of the studies stated that the interval between the index and reference tests may have caused deterioration of sample quality, but most ensured that samples were frozen if the reference test was not performed on-site.

In terms of applicability concerns, most (90%) of the studies were aligned with the current study’s research question.

Diagnostic performance

Forest plots for the included index tests with two or more studies are presented in Figure 4, allowing a visual comparison of each index test group. The sensitivity and specificity measures for each study in an index test group are plotted with 95% confidence intervals. The results of the final meta-analysis are presented in Table 3 as an overall summary point for the sensitivity and specificity of each included index test. Bivariate modelling was performed for POC-CCA, portable/mobile-phone microscopy, and LAMP, as these tests use the same threshold for determining positivity. HSROC modelling was used for the RPA test, as multiple thresholds were used for determining positivity in the included studies.

Table 3. Pooled sensitivity and specificity results for rapid and POC index tests that diagnose urogenital schistosomiasis for individuals infected in African settings.

Index testtypeNo. of accuracy recordsPooled number of participantsNo. with schistosomiasis (%)Sensitivity %(95% CI)Specificity %(95% CI)
Serological tests
IHA2198139 (70.2)59.5% (35.8-79.5)76.1% (65.1-84.5)
ICT-RDT2804219 (27.2)98.6% (77.0-99.9)47.7% (27.7-69.5)
RDT-Sh115078 (52.0)90.0% (81.0-95.0)54.0% (42.0-65.0)
SmCTF-RDT220977 (36.8)98.4% (7.70-99.9)36.7% (27.4-47.6)
DDIA114689 (60.9)60.0% (49.0-70.0)61.0% (48.0-74.0)
Antigen Tests
POC-CCA83751887 (23.6)49.8% (25.4-73.4)77.6% (65.1-86.5)
NAAT tests
LAMP5542230 (42.4)94.7% (67.2-99.3)66.4% (46.99-81.6)
RPA3830235 (28.3)94.4% (90.0-96.0)96.6% (96.0-99.0)
Parasitological tests
Portable or mobile microscopy122497805 (32.2)81.5% (72.6-88.0)91.4% (86.5-94.6)

IHA: Indirect haemagglutination assay, ICT-RDT: Immunochromatographic rapid diagnostic test, SmCTF-RDT: S. mansoni-cercarial transformation fluid- RDT, DDIA: Dipstick dye immunoassay, RDT-Sh: Rapid Diagnostic Test- S. haematobium, POC-CCA: Point of Care- Circulating Cathodic antigen, LAMP: Loop Mediated Isothermal Amplification, RPA: Recombinase Polymerase Amplification.

In the final meta-analysis, the molecular test RPA had the highest overall accuracy (sensitivity, 94.4%; specificity, 96.6%), displaying the best trade-off between sensitivity and specificity. Based on this meta-analysis, portable/mobile-phone microscopy (sensitivity, 81.5%; specificity, 91.4%) had the second-highest diagnostic accuracy. LAMP (sensitivity, 94.7%; specificity, 66.4%) and the serological screening test ICT-RDT (sensitivity, 98.6%; specificity, 47.7%) come close in accuracy for S. haematobium diagnosis, where both are sensitive tests but lack specificity. Tests with moderate accuracy are the SmCTF-RDT and RDT-Sh serological tests, where further research with improved reference tests could provide answers for their use as rapid tests. Our overall results indicate that the IHA (sensitivity, 59.5%; specificity, 76.1%), DDIA (sensitivity, 60.0%; specificity, 61.0%), and POC-CCA (sensitivity, 48.9%; specificity, 77.3%) index tests have low diagnostic accuracy for S. haematobium and may not be suitable for prevalence mapping or estimation of infection. These tests miss infections that are identified by laboratory-based tests and are therefore not useful in practice.

Summary ROC curve for all included index tests

To enable comparisons, the summary ROC curves for all index tests are presented in a single graph in Figure 3. Tests with higher accuracy are in the top left corner. This shows that the molecular test RPA has the highest overall accuracy. Portable/mobile-phone microscopy, ICT-RDT, and LAMP are grouped as the second-highest diagnostic accuracy, while POC-CCA and DDIA have the poorest diagnostic accuracy.

Summary ROC curves for all included index tests.
Figure 3. Summary ROC curves for all included index tests.
Forest Plot of included studies grouped according to the index test.
Figure 4. Forest Plot of included studies grouped according to the index test.

Exploration of heterogeneity

The forest plots (Figure 4) and the individual SROC curves for each index test group (Supplementary Additional File 3) show visually that there is high heterogeneity among the studies for POC-CCA and LAMP, and moderate heterogeneity for RPA and portable/mobile-phone microscopes. To further determine possible reasons for heterogeneity, an exploratory analysis using forest plots and SROC plots (Supplementary Additional File 4) and a subgroup meta-analysis (Table 4) were performed for the covariates: prevalence setting, reference standard, and age group.

Table 4. Subgroup analysis for individual index tests comparing overall accuracy results with subgroup accuracy results.

Co-variateSub-group (records)Participants(n)Sensitivity (%) (95% CI)Specificity (%) (95% CI)
Antigen tests
POC-CCA: Overall accuracy (8 records)N=375149.8 (25.4-73.4)77.6 (65.1-86.5)
Prevalence setting highEndemic high Sh prevalent (3)67618.8 (0.0- 91.7)59.9 (43.4-74.4)
Endemic high co-endemic Sh/Sm (1)50067.0 (59.0-74.0)83.0 (76.0-88.0)
Prevalence setting lowEndemic low Sh prevalent (2)205072.6 (34.7-92.9)87.6 (72.3-95.0)
Endemic low co-endemic Sh/Sm (2)52545.6 (36.0-55.5)84.3 (70.0-92.5)
Age groupAdults>18 years (1)50067.0 (59.0-74.0)83.0 (76.0-88.0)
Children (7)325145.9 (19.8-7.5)76.8 (62.-86.9)
NAAT tests
LAMP: Overall accuracy (5 records)(N = 542)94.7 (67.2-99.3)66.4 (46.99-81.6)
Reference standardPCR (2)21194.3 (50.0-99.6)73.7 (41.5-91.7)
Single microscopy (3)33191.4 (64.5-98.4)50.7 (39.0-62.4)
Age groupAdults& children (3)28098.0 (47.6-99.9)70.5 (46.7-86.7)
School Children only (2)26282.4 (74.1-88.4)45.2 (35.3-55.5)
Prevalence settingEndemic high Sh prevalent (3)46979.0 (69.3-86.3)59.6 (35.6-79.8)
Endemic high co-endemic Sh/Sm (2)15599.4 (46.6-99.9)63.8 (36.9-84.2)
Parasitological tests
Portable microscopy: Overall accuracy (12 records)(N = 2497)81.5 (72.6-88.0)91.4 (86.5-94.6)
Device typePortable microscope (2)Cellscope (2)Schistoscope (8)449275177376.5 (68.2-82.3)59.0 (47.3-73.5)86.3 (77.7-91.9)98.5 (97.5-99.2)99.5 (97.5-99.2)88.1 (82.8-92.0)

Sh, S. haematobium; Sm, S. mansoni.

For POC-CCA, the low-prevalence setting in an S. haematobium-prevalent area had improved sensitivity (72.6% vs 49.8%), and poor sensitivity results were found for high Sh-prevalent areas (18.8%). A high S. haematobium prevalence is therefore not suitable for diagnosis using the POC-CAA test due to many false positives, and possibly due to cross reactions with S. mansoni. Single microscopy was used as a reference for all the POC-CCA studies, which may have caused bias in the specificity results of 3 of the POC-CCA studies that demonstrate low Sh prevalence.

For LAMP, the quality of the reference test affects the quality of the results: specificity improves (73.7% vs 66.4%) when PCR is used as the reference test, although sensitivity does not change much. The PCR reference standards show superior diagnostic performance compared to single microscopy, implying that accuracy is inaccurately portrayed by the usual gold standard using single microscopy of filtered urine.

Age group also caused heterogeneity in the results for LAMP and POC-CCA where a combined population of all ages had a better sensitivity than school-aged children (98.0% vs 82.4%) and (67.0% vs 45.9%), respectively.

The Schistoscope revealed the highest sensitivity (86.3%) of the three portable microscopes, where the Cellscope and mobile-phone attached microscope had lower sensitivities (59.0% and 76.5%, respectively), but improved specificities. The studies were all performed in high-moderate endemic settings, and further validation could determine the accuracy of portable microscopes in low-prevalence settings.

Discussion

This systematic review sought to identify a rapid test with high sensitivity and specificity to improve prevalence mapping and elimination-of-transmission testing, replacing microscopy in integrated control strategies in low-resource and rural areas. The results of this review can support the choice of diagnostic test in endemic areas with poor laboratory infrastructure and collate the available evidence on the accuracy measures of rapid field-based diagnostic tests for the detection of S. haematobium. Previous systematic and narrative reviews have focused on testing and screening both S. mansoni and S. haematobium (Vaillant et al., 2024) (Rivera et al., 2024) and not specifically for field-based testing in low-resource areas. Recent reviews have been published focusing on individual tests, for example, the LAMP assay in low-endemic areas (Zhou et al., 2025), and AI-assisted diagnostic tools (Desale et al., 2026). This review provides a broader synthesis of multiple rapid and field-deployable diagnostic tools in low-resource areas and thereby offers a comparison between the performance of various rapid field-based tests.

A key finding was that the field-based molecular tests RPA and LAMP are the most suitable tests for prevalence estimation of active infections in both high and low prevalence settings and for various stages of chronicity. RPA has the highest accuracy in low-intensity and chronic infections and is also most useful for the estimation of the elimination of transmission of S. haematobium. These tests are still in development and require portable testing devices to obtain results from urine samples but show great promise for use as test-and-treat tools in low-resource areas with field laboratories. Considering the REASSURED criteria (Land et al., 2019) as the benchmark to assess POC and rapid tests, the RPA and LAMP offer an affordable, sensitive, and specific solution.

The portable/mobile-phone microscopy tests using handheld cell phone devices and AI-assisted microscopy, such as the Schistoscope (sensitivity, 81.4%, specificity, 91.4%) have great value in areas where skilled microscopists are unavailable and show promise in the automation of egg detection using microscopy of urine.

Another finding was that the ICT-RDT and SmCTF-RDT serological tests have high sensitivity (98.6% and 98.4%, respectively) in all prevalence settings. However, these tests require trained phlebotomists to draw blood and portable centrifuges to separate serum from whole blood. The SmCTF-RDT can also be performed using finger-prick blood, which is far easier to obtain. The low specificity of the ICT-RDT and SmCTF-RDT tests (47.7% and 36.7%, respectively) was due to reference test bias, as the comparator tests were microscopy in a low-prevalence area. These serological tests do not differentiate between active or past infection, and between S. haematobium and S mansoni. They are therefore excellent screening tests for travellers and migrants and in areas of co-endemicity, detecting infections with either species of schistosomiasis. Both species are treated with Praziquantel, which can be administered in the early stage of infection (Zwang and Olliaro, 2014).

Our review also determined that two tests have low sensitivity and, therefore lower diagnostic accuracy than conventional microscopy. These are IHA (sensitivity, 59.5%, specificity, 76.1%), and DDIA serological test (sensitivity, 60.0%, specificity, 61.0%). These tests are moderately sensitive and therefore not suitable for prevalence mapping or estimation of infection.

The findings of this study agree with those of previous reviews (Ochodo et al., 2015) (Stothard et al., 2006) that the POC-CCA lateral flow test, used extensively for S. mansoni diagnosis in Africa, has low diagnostic accuracy for S. haematobium (sensitivity, 49.8%, specificity, 77.6%). The test results in false negatives and misses many S. haematobium infections detected using laboratory-based tests. There was a high heterogeneity for POC-CCA results, where the S. haematobium-prevalent areas had low sensitivity, even if infection prevalence is high. As the POC-CCA is not species-specific, the test should be used with caution in co-endemic areas, needing secondary testing using laboratory PCR or urinary filtration with microscopy to confirm results (Lorenz et al., 2025). The POC-CCA test, therefore, appears not to be useful in practice or for prevalence screening for S. haematobium. It is, however, a sensitive screening tool in endemic areas for intestinal schistosomiasis caused by S. mansoni (sensitivity 93.4%, specificity 63.5%) (Alemu et al., 2025).

The most promising studies use improved reference standards such as PCR, double/triple microscopy, or composite tests. Microscopy is an imperfect reference standard that introduces bias for tests that are more sensitive than microscopy. Five of the nine tests eligible for this review were found to have a high false positive rate, which indicates that the test detects more infections than the reference test. Therefore, specificity results for LAMP, SmCTF-RDT, and ICT-RDT are possibly biased and need to be verified in future accuracy studies that use a higher quality reference test. Future test performance studies in low-prevalence areas should use higher-quality reference tests instead of single microscopy to improve and verify the accuracy of the test performance results of this review.

Study limitations

A major strength of this review was the robust methodological approach, considering the challenges in diagnosing schistosomiasis. However, meta-analyses require two or more studies with a similar experimental design to enable accurate comparison. In this review, four or less highly heterogeneous studies were available for six of the nine index tests analysed. This resulted in low certainty for most of the tests evaluated in this review. We were able to conduct an adequate meta-analysis for POC-CCA, LAMP, and portable microscope tests, as there were more than four studies for each. Reference test bias occurred in more than 40% of the studies, where single microscopy was used as the reference test in areas with an infection prevalence of <50% among school-aged children, resulting in low specificity results. The specificity results may therefore not be accurate for all nine of the tests in this review.

Conclusion

Diagnostic gaps have been recognized in areas in Africa where S. haematobium is the predominant species of schistosomiasis infection. This review has identified possible tests that have the potential to replace conventional microscopy, particularly in low-resource areas where patients have limited access to health facilities. The rapidity of a test, and its result at the patient’s side is of great value to patients who otherwise may not return readily to receive their results, especially in areas with few or poorly equipped health facilities. We have identified two molecular tests, RPA and LAMP, that can be implemented to diagnose acute, sub-acute, and chronic stages of urogenital schistosomiasis in field/mobile laboratories in poor resource areas with some training of laboratory staff and minimal equipment. We have also identified portable and AI-Assisted microscopes as highly sensitive and rapid screening tests for S. haematobium in endemic areas of high prevalence/infection intensity of >50%. Two serological lateral flow screening tests ICT-RDT and SmCTF-RDT are highly sensitive tests able to diagnose all species of Schistosoma, and effective for mapping and prevalence screening with rapid results of two hours or less. Both tests have a low specificity result, which may be compromised by reference test bias, and need further testing to confirm their diagnostic accuracy in endemic areas.

An effective control program should implement testing of populations after mass drug administration to determine whether transmission interruption has occurred in a particular area. We have identified RPA as a highly accurate rapid test that functions optimally in low-prevalence settings of up to 2.5% infection prevalence. For the implementation of successful control programs in endemic areas, a test-and-treat strategy may be of great value for improving the support of parents, teachers, and community leaders in gaining the consent of the recommended 75% of a community for mass drug administration strategies. Such programs, however, need to become routine and sustained. Furthermore, identifying and treating patients with chronic or low levels of infection is of paramount value to prevent further transmission of this disease and to prevent the debilitating, long-term morbidities associated with untreated urinary schistosomiasis.

Correction note

A correction has been made to this article. Details can be found at: 10.3389/fpara.2026.1960349.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fpara.2026.1830789/full#supplementary-material

Supplementary material Supplementary material at the source

Footnotes

Edited by: Lisette Van Lieshout, Leiden University Medical Center (LUMC), Netherlands

Reviewed by: Cristina Toscano Fonseca, Fundação Oswaldo Cruz., Brazil

Pytsje T. Hoekstra, Leiden University Medical Center (LUMC), Netherlands

Statements

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Funding

The author(s) declared that financial support was received for this work and/or its publication. The research leading to these results has been partially funded by Horizon Europe, Grant Agreement no. 101057853 (DUALSAVE-FGS). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.

Data availability

The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding author.

Author contributions

ED: Visualization, Software, Conceptualization, Investigation, Funding acquisition, Writing – original draft, Resources, Writing – review & editing, Formal analysis, Validation, Project administration, Methodology, Data curation. EO: Formal analysis, Supervision, Methodology, Software, Conceptualization, Writing – review & editing. PP: Supervision, Writing – review & editing, Validation, Conceptualization. NN: Conceptualization, Supervision, Writing – review & editing, Formal analysis, Validation, Project administration.

Acknowledgements

Daniel Eklapa (MSc Wits) as secondary reviewer. Holijah Uy (PhD UCT) for input on RevMan software.

References

  1. Adenowo A. F., Oyinloye B. E., Ogunyinka B. I., Kappo A. P. (2015). Impact of human schistosomiasis in sub-Saharan Africa. Braz. J. Infect. Dis. 19, 196 – 205. doi: doi:10.1016/j.bjid.2014.11.004
  2. Alemu G., Nibret E., Munshea A., Anegagrie M., Flores-Chávez M., Hailu T.. (2025). Comparative performance of Kato–Katz, POC-CCA and real-time PCR in detecting Schistosoma mansoni infection at different endemicity settings in northwest Ethiopia: a cross-sectional study. Trop. Med. Health 53, 103. doi: doi:10.1186/s41182-025-00777-7
  3. Ally O., Kanoi B. N., Ochola L., Nyanjom S. G., Shiluli C., Misinzo G.. (2024). Schistosomiasis diagnosis: Challenges and opportunities for elimination. PloS Negl.Trop. Dis. 18, e0012282. doi: doi:10.1371/journal.pntd.0012282
  4. Archer J., Barksby R., Pennance T., Rostron P., Bakar F., Knopp S.. (2020). Analytical and clinical assessment of a portable, isothermal recombinase polymerase amplification (RPA) assay for the molecular diagnosis of urogenital schistosomiasis. Molecules 25, 4175. doi: doi:10.3390/molecules25184175
  5. Archer J. I., Patwary F. K., Sturt I. A. S., Webb E. L., Rutty Phiri C., Mweene T.. (2022). Validation of the isothermal Schistosoma haematobium recombinase polymerase amplification (RPA) assay, coupled with simplified sample preparation, for diagnosing female genital schistosomiasis using cervicovaginal lavage and vaginal self-swab samples. PLoS Negl. Trop. Dis. 16. doi: doi:10.1371/journal.pntd.0010276
  6. Armstrong M., Harris A. R., D’Ambrosio M. V., Coulibaly J. T., Essien-Baidoo S., Ephraim R. K. D.. (2022). Point-of-care sample preparation and automated quantitative detection of Schistosoma haematobium using mobile phone microscopy. Am. J. Trop. Med. Hyg. 106, 1442 – 1449. doi: doi:10.4269/ajtmh.21-1071
  7. Ashton R. A., Stewart B. T., Petty N., Lado M., Finn T., Brooker S.. (2011). Accuracy of circulating cathodic antigen tests for rapid mapping of Schistosoma mansoni and S. haematobium infections in Southern Sudan. Trop. Med. Int. Health 16, 1099 – 1103. doi: doi:10.1111/j.1365-3156.2011.02815.x
  8. Ayele B., Erko B., Legesse M., Hailu A., Medhin G. (2008). Evaluation of circulating cathodic antigen (CCA) strip for diagnosis of urinary schistosomiasis in Hassoba school children, Afar, Ethiopia. Parasite 15, 69 – 75. doi: doi:10.1051/parasite/2008151069
  9. Bayoumi A., Al-Refai S., Badr M., Abd El-Aal A. (2016). Loop-mediated isothermal amplification (LAMP): Sensitive and rapid detection of Schistosoma haematobium DNA in urine samples of Egyptian suspected cases. J. Egyptian Soc. Parisitology 46, 299 – 308. doi: doi:10.21608/jesp.2016.88677
  10. Bogoch I. I., Koydemir H. C., Tseng D., Ephraim R. K. D., Duah E., Tee J.. (2017). Evaluation of a mobile phone-based microscope for screening of Schistosoma haematobium infection in rural Ghana. Am. J. Trop. Med. Hyg. 96, 1468 – 1471. doi: doi:10.4269/ajtmh.16-0912
  11. Chala B. (2023). Advances in diagnosis of schistosomiasis: Focus on challenges and future approaches. Int. J. Gen. Med. 16, 983 – 995. doi: doi:10.2147/IJGM.S391017
  12. Coulibaly J. T., N’Goran E. K., Utzinger J., Doenhoff M. J., Dawson E. M. (2013). A new rapid diagnostic test for detection of anti-Schistosoma mansoni and anti-Schistosoma haematobium antibodies. Parasitol. Vectors 6, 29. doi: doi:10.1186/1756-3305-6-29
  13. Coulibaly J. T., Ouattara M., D’Ambrosio M. V., Fletcher D. A., Keiser J., Utzinger J.. (2016). Accuracy of mobile phone and handheld light microscopy for the diagnosis of schistosomiasis and intestinal protozoa infections in Côte d’Ivoire. PloS Negl.Trop. Dis. 10. doi: doi:10.1371/journal.pntd.0004768
  14. Coulibaly J. T., Silue K. D., Armstrong M., de León Derby M. D., D’Ambrosio M. V., Fletcher D. A.. (2023). High sensitivity of mobile phone microscopy screening for Schistosoma haematobium in Azaguié, Côte d’Ivoire. Am. J. Trop. Med. Hyg. 108, 41 – 43. doi: doi:10.4269/ajtmh.22-0527
  15. Danso-Appiah A., Minton J., Boamah D., Otchere J., Asmah R. H., Rodgers M.. (2016). Accuracy of point-of-care testing for circulatory cathodic antigen in the detection of schistosome infection: systematic review and meta-analysis. Bull. World Health Organ. 94, 522A – 533A. doi: doi:10.2471/BLT.15.158741
  16. Desale S., Alemu G., Hailu T. (2026). Accuracy of AI-assisted diagnostic tools for Schistosoma haematobium: A systematic review and meta-analysis. PloS Negl.Trop. Dis. 20, e0013703. doi: doi:10.1371/journal.pntd.0013703
  17. Dinnes J., Deeks J. J., Leeflang M. M., Li T. (2022). Chapter 7: collecting data. In: Deeks J. J., Bossuyt P. M., Leeflang M. M., Takwoingi Y., editors. Cochrane Handbook for Systematic Reviews of Diagnostic Test Accuracy. (London: Cochrane). doi: doi:10.1002/9781119756194.ch7
  18. Drain P. K., Hyle E. P., Noubary F., Freedberg K. A., Wilson D., Bishai W. R.. (2014). Diagnostic point-of-care tests in resource-limited settings. Lancet Infect. Dis. 14, 239 – 249. doi: doi:10.1016/S1473-3099(13)70250-0
  19. El-Ghareeb A. S., Abd El Motaleb G. S., Waked N. M., Osman Hany Kamel N., Aly N. S. (2016). Circulating cathodic antigen cassette test versus haematuria strip test in diagnosis of urinary schistosomiasis. J. Parasitic Dis. 40, 1193 – 1198. doi: doi:10.1007/s12639-015-0648-2
  20. Ephraim R. K. D., Cybulski J. S., Duah E., Prakash M., D’Ambrosio M. V., Fletcher D. A.. (2015). Diagnosis of Schistosoma haematobium infection with a mobile phone-mounted Foldscope and a reversed-lens CellScope in Ghana. Am. J. Trop. Med. Hyg. 92, 1253 – 1256. doi: doi:10.4269/ajtmh.14-0741
  21. Feleke D. G., Alemu Y., Bisetegn H., Debash H. (2023). Accuracy of diagnostic tests for detecting Schistosoma mansoni and S. haematobium in sub-Saharan Africa: A systematic review and meta-analysis. BioMed. Res. Int. 2023, 1 – 13. doi: doi:10.1155/2023/3769931
  22. Frimpong M., Kyei-Tuffuor L., Fondjo L. A., Ahor H. S., Adjei-Kusi P., Maiga-Ascofare O.. (2021). Evaluation of a real-time recombinase polymerase amplification assay for rapid detection of Schistosoma haematobium infection in resource-limited setting. Acta Trop. 216. doi: doi:10.1016/j.actatropica.2021.105847
  23. Gandasegui J., Fernández-Soto P., Carranza-Rodríguez C., Pérez-Arellano J. L., Vicente B., López-Abán J.. (2015). The rapid-heat LAMPellet method: A potential diagnostic method for human urogenital schistosomiasis. PloS Negl.Trop. Dis. 9, 1 – 23. doi: doi:10.1371/journal.pntd.0003963
  24. Gandasegui J., Fernández-Soto P., Dacal E., Rodríguez E., Saugar J. M., Yepes E.. (2018). Field and laboratory comparative evaluation of a LAMP assay for the diagnosis of urogenital schistosomiasis in Cubal, Central Angola. Trop. Med. Int. Health 23, 992 – 1001. doi: doi:10.1111/tmi.13117
  25. González-Sanz M., Martín-Rubio I., Martín O., Muriel A., de la Fuente-Hernanz S., Crespillo-Andújar C.. (2025). Description of the serological response after treatment of chronic imported schistosomiasis. Trop. Med. Infect. Dis. 10, 22. doi: doi:10.3390/tropicalmed10010022
  26. Gray D. J., Ross A. G., Li Y. S., McManus D. P. (2011). Diagnosis and management of schistosomiasis. BMJ 342, d2651. doi: doi:10.1136/bmj.d2651
  27. Harbord R. M., Whiting P. (2009). metandi: Meta-analysis of diagnostic accuracy using hierarchical logistic regression. Stata J. 9, 211 – 229. doi: doi:10.1177/1536867x0900900203
  28. Hong S. T. (2023). Review of recent prevalence of urogenital schistosomiasis in sub-Saharan Africa and diagnostic challenges in the field setting 13, 1680. doi: doi:10.3390/life13081670
  29. Houmsou R. S., Wama B. E., Agere H., Uniga J. A., Jerry T. J., Azuaga P.. (2021). Diagnostic accuracy of Schistosoma ICT IgG-IgM and comparison to other used techniques screening urinary schistosomiasis in Nigeria. Adv. Lab. Med. 2, 71 – 77. doi: doi:10.1515/almed-2020-0093
  30. Land K. J., Boeras D. I., Chen X. S., Ramsay A. R., Peeling R. W. (2019). REASSURED diagnostics to inform disease control strategies, strengthen health systems and improve patient outcomes. Nat. Microbiol. 4, 46 – 54. doi: doi:10.1038/s41564-018-0295-3
  31. Li Q., Li Y. L., Guo S. Y., Li S. Z., Wang Q., Lin W. N.. (2025). Global trends of schistosomiasis burden from 1990 to 2021 across 204 countries and territories: Findings from GBD 2021 study. Acta Trop. 261, 107504. doi: doi:10.1016/j.actatropica.2024.107504
  32. Lodh N., Mikita K., Bosompem K. M., Anyan W. K., Quartey J. K., Otchere J.. (2017). Point of care diagnosis of multiple schistosome parasites: Species-specific DNA detection in urine by loop-mediated isothermal amplification (LAMP). Acta Trop. 173, 125 – 129. doi: doi:10.1016/j.actatropica.2017.06.015
  33. Lorenz E., Razafindrakoto R., Rausche P., Rasolojaona Z. T., Razafindralava N. M., Zerbo A.. (2025). Detecting Schistosoma infections in endemic countries: a diagnostic accuracy study in rural Madagascar. Infect. Dis. Poverty 14, 20. doi: doi:10.1186/s40249-025-01292-x
  34. Mabey D., Peeling R. W., Ustianowski A., Perkins M. D. (2004). Diagnostics for the developing world. Nat. Rev. Microbiol. 2, 231 – 240. doi: doi:10.1038/nrmicro841
  35. Macaskill P., Gatsonis C., Deeks J., Harbord R., Takwoingi Y. (2010). “ Cochrane handbook for systematic reviews of diagnostic test accuracy chapter 10 analysing and presenting results,” in Cochrane Handbook for Systematic Reviews of Diagnostic Test Accuracy Version 1.0. The Cochrane Collaboration. Deeks J. J.. Available online at: http://srdta.cochrane.org/.
  36. Mcinnes M. D. F., Moher D., Thombs B. D., Mcgrath T. A., Bossuyt P. M. (2018). Preferred reporting items for a systematic review and meta-analysis of diagnostic test accuracy studies the PRISMA-DTA statement supplemental content CME quiz at jamanetwork.com/learning. JAMA 319, 388 – 396. doi: doi:10.1001/jama.2017.19163
  37. Meulah B., Oyibo P., Bengtson M., Agbana T., Lontchi R. A. L., Adegnika A. A.. (2022). Performance evaluation of the Schistoscope 5.0 for (semi-)automated digital detection and quantification of Schistosoma haematobium eggs in urine: A field-based study in Nigeria. Am. J. Trop. Med. Hyg. 107, 1047 – 1054. doi: doi:10.4269/ajtmh.22-0276
  38. Meulah B., Oyibo P., Hoekstra P. T., Nguema Moure P. A., Nzamba Maloum M., Laclong-Lontchi R. A.. (2024). Validation of artificial intelligence-based digital microscopy for automated detection of Schistosoma haematobium eggs in urine in Gabon. PloS Negl.Trop. Dis. 18, e0011967. doi: doi:10.1371/journal.pntd.0011967
  39. Mudenda J., Hamooya B. M., Tembo S., Halwindi H., Siwila J., Phiri M. M. (2022). Diagnostic accuracy of Schistosoma immunochromatographic IgG/IgM rapid test in the detection of schistosomiasis in Zambia. J. Basic Appl. Zool. 83. doi: doi:10.1186/s41936-022-00265-y
  40. Naaktgeboren C. A., Ochodo E. A., Van Enst W. A., De Groot J. A. H., Hooft L., Leeflang M. M. G.. (2016). Assessing variability in results in systematic reviews of diagnostic studies. BMC Med. Res. Methodol. 16, 6. doi: doi:10.1186/s12874-016-0108-4
  41. Nausch N., Dawson E. M., Midzi N., Mduluza T., Mutapi F., Doenhoff M. J. (2014). Field evaluation of a new antibody-based diagnostic for Schistosoma haematobium and S. mansoni at the point-of-care in northeast Zimbabwe. BMC Infect. Dis. 14, 165. doi: doi:10.1186/1471-2334-14-165
  42. Obeng B. B., Aryeetey Y. A., De Dood C. J., Amoah A. S., Larbi I. A., Deelder A. M.. (2008). Application of a circulating-cathodic-antigen (CCA) strip test and real-time PCR, in comparison with microscopy, for the detection of Schistosoma haematobium in urine samples from Ghana. Ann. Trop. Med. Parasitol. 102, 625 – 633. doi: doi:10.1179/136485908X337490
  43. Ochodo E. A., Gopalakrishna G., Spek B., Reitsma J. B., van Lieshout L., Polman K.. (2015). Circulating antigen tests and urine reagent strips for diagnosis of active schistosomiasis in endemic areas. Cochrane Database Systematic Rev. 2015. doi: doi:10.1002/14651858.CD009579.pub2
  44. Poggensee G., Kiwelu I., Saria M., Richter J., Krantz I., Feldmeier H. (1998). Schistosomiasis of the lower reproductive tract without egg excretion in urine. Am. J. Trop. Med. Hyg. 59, 782 – 783. doi: doi:10.4269/ajtmh.1998.59.782
  45. Randjelovic A., Frønæs S. G., Munsami M., Kvalsvig J. D., Zulu S. G., Gagai S.. (2015). A study of hurdles in mass treatment of schistosomiasis in Kwazulu-Natal, South Africa. South. Afr. Family Pract. 57, 57 – 61. doi: doi:10.1080/20786190.2014.978121
  46. Reitsma J. B., Glas A. S., Rutjes A. W. S., Scholten R. J. P. M., Bossuyt P. M., Zwinderman A. H. (2005). Bivariate analysis of sensitivity and specificity produces informative summary measures in diagnostic reviews. J. Clin. Epidemiol. 58, 982 – 990. doi: doi:10.1016/j.jclinepi.2005.02.022
  47. Rivera J., Mu Y., Gordon C. A., Jones M. K. (2024). Current and upcoming point-of-care diagnostics for schistosomiasis. Trends Parasitol. 40, 60 – 74. doi: doi:10.1016/j.pt.2023.10.005
  48. Robinson K. E., Grewal E. P., Pritt B. S., Lloyd M., Stephano A. M., Fardine M.. (2021). Schistosomiasis prevalence and low-cost diagnostics in rural Northwestern Madagascar: A pilot study. J. Glob. Health Rep. 5, e2021034. doi: doi:10.29392/001c.22123
  49. Rubaba O., Chimbari M. J., Soko W., Manyangadze T., Mukaratirwa S. (2018). Validation of a urine circulating cathodic antigen cassette test for detection of Schistosoma haematobium in uMkhanyakude district of South Africa. Acta Trop. 182, 161 – 165. doi: doi:10.1016/j.actatropica.2018.02.029
  50. Rutter C. M., Gatsonis C. A. (2001). A hierarchical regression approach to meta-analysis of diagnostic test accuracy evaluations. Stat. Med. 20, 2865 – 2884. doi: doi:10.1002/sim.942
  51. Sanneh B., Joof E., Sanyang A. M., Renneker K., Camara Y., Sey A. P.. (2017). Field evaluation of a schistosome circulating cathodic antigen rapid test kit at point-of-care for mapping of schistosomiasis endemic districts in The Gambia. PloS One 12. doi: doi:10.1371/journal.pone.0182003
  52. Sheele J. M., Kihara J. H., Baddorf S., Byrne J., Ravi B. (2013). Evaluation of a novel rapid diagnostic test for Schistosoma haematobium based on the detection of human immunoglobulins bound to filtered Schistosoma haematobium eggs. Trop. Med. Int. Health 18, 477 – 484. doi: doi:10.1111/tmi.12063
  53. Shen Z., Luo H. (2025). The impact of schistosomiasis on the Global Disease Burden: a systematic analysis based on the 2021 Global Burden of Disease study. Parasite 32, 12. doi: doi:10.1051/parasite/2025005
  54. Stothard J. R., Chitsulo L., Kristensen T. K., Utzinger J. (2009). Control of schistosomiasis in sub-Saharan Africa: Progress made, new opportunities and remaining challenges. Parasitology 136, 1665 – 1675. doi: doi:10.1017/S0031182009991272
  55. Stothard J. R., Kabatereine N. B., Tukahebwa E. M., Kazibwe F., Rollinson D., Mathieson W.. (2006). Use of circulating cathodic antigen (CCA) dipsticks for detection of intestinal and urinary schistosomiasis. Acta Trop. 97, 219 – 228. doi: doi:10.1016/j.actatropica.2005.11.004
  56. Sturt A. S., Webb E. L., Francis S. C., Hayes R. J., Bustinduy A. L. (2020). Beyond the barrier: Female Genital Schistosomiasis as a potential risk factor for HIV-1 acquisition. Acta Trop. 209, 105524. doi: doi:10.1016/j.actatropica.2020.105524
  57. Vaillant M. T., Philippy F., Neven A., Barré J., Bulaev D., Olliaro P. L.. (2024). Diagnostic tests for human Schistosoma mansoni and Schistosoma haematobium infection: a systematic review and meta-analysis. Lancet Microbe 5, e366 – e378. doi: doi:10.1016/S2666-5247(23)00377-4
  58. van Bergen K. J. M., Brienen E. A. T., Randrianasolo B. S., Ramarokoto C. E., Leutscher P., Kjetland E. F.. (2024). Next step towards point-of-care molecular diagnosis of female genital schistosomiasis (FGS): evaluation of an instrument-free LAMP procedure. Front. Parasitol. 3. doi: doi:10.3389/fpara.2024.1297310
  59. Wami W. M., Nausch N., Bauer K., Midzi N., Gwisai R., Simmonds P.. (2014). Comparing parasitological vs serological determination of Schistosoma haematobium infection prevalence in preschool and primary school-aged children: Implications for control programmes. Parasitology 141, 1962 – 1970. doi: doi:10.1017/S0031182014000213
  60. Whiting P. F., Rutjes A. W. S., Westwood M. E., Mallett S., Deeks J. J., Reitsma J. B.. (2011). QUADAS-2: A revised tool for the quality assessment of diagnostic accuracy studies. Ann. Int. Med. 155, 529 – 536. doi: doi:10.7326/0003-4819-155-8-201110180-00009
  61. World Health Organization (2014). In Vitro Diagnostics: Simple/ Rapid Tests [Internet]. Available online at: https://www.who.int/news-room/questions-and-answers/item/simple-rapid-tests.
  62. World Health Organization (2022). WHO Guideline on Control and Elimination of Human Schistosomiasis (Geneva: World Health Organization).
  63. Yameny A. (2019). The validity of indirect haemagglutination assay (IHA) in the detection of Schistosoma haematobium infection relative to microscopic examination. J. Med. Life. Sci. 1, 57 – 64. doi: doi:10.21608/jmals.2019.68364
  64. Zhang L. J., Mwanakasale V., Xu J., Sun L. P., Yin X. M., Zhang J. F.. (2020). Diagnostic performance of two specific schistosoma japonicum immunological tests for screening schistosoma haematobium in school children in Zambia. Acta Trop. 202, 105285. doi: doi:10.1016/j.actatropica.2019.105285
  65. Zhou X., Li J., Qiu J., Feng T., Lv C., Deng W.. (2025). Test accuracy of loop-mediated isothermal amplification for schistosomiasis in low endemicity areas: a systematic review and meta-analysis. Infect. Dis. Poverty 14. doi: doi:10.1186/s40249-025-01346-0
  66. Zwang J., Olliaro P. L. (2014). Clinical efficacy and tolerability of praziquantel for intestinal and urinary schistosomiasis—A meta-analysis of comparative and non-comparative clinical trials. PloS Negl.Trop. Dis. 8, e3286. doi: doi:10.1371/journal.pntd.0003286

Citation

Draxl, E., et al. (2026). Diagnostic accuracy of rapid and point-of-care tests for Schistosoma haematobium in African populations: a systematic review and meta-analysis. Frontiers in Parasitology, Volume 5. https://doi.org/
Received
Accepted
Published
Volume
5 - 2026

Copyright

Copyright © 2026 Draxl, Ochodo, Pillay and Ndlovu.

License: https://creativecommons.org/licenses/by/4.0/

*Correspondence: Erica Draxl

Article metrics

Views
0
Downloads
0
Citations
0
Shares
0